23
METHOD PROCEDURE
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add one packet of
HI93750B-0
Potassium Reagent B. Replace the
plastic stopper and the cap. Shake
gently for 1 minute.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Read
and the display will show a 3 minutes
countdown prior to the measurement. To skip the timer,
press
Read
twice.
• After the 3 minutes countdown has passed, invert the
cuvette 5 times to mix.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Read
to start reading. The instrument displays the results in
mg/L
of
potassium (K).