22
METHOD PROCEDURE
8. METHOD PROCEDURE
8.1. POTASSIUM LOW RANGE
REQUIRED REAGENTS
Code Description Quantity
HI93750A-0
Potassium Reagent A
6 drops
HI93750B-0
Potassium Reagent B
1 packet
REAGENT SETS
HI93750-01
Potassium Reagent - 100 tests
HI93750-03
Potassium Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
Potassium LR
using the procedure described in the RANGE SELECTION
section.
Note:
If tutorial mode is disabled follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of sample (up to
the mark).
• Add 6 drops of
HI93750A-0
Potassium
Reagent A. Replace the plastic stopper and the
cap. Swirl the solution.
• Insert the cuvette into the holder and ensure
that the notch on the cap is positioned securely
in the groove.
10 mL
×
6