6.  GENERAL OPERATIONS

6.1.  METER VALIDATION: CAL CHECK & CALIBRATION

Validation of the 

HI97712

 involves verifying the concentration of the certified CAL Check 

standards. The CAL Check screen guides the user step-by-step through the validation 
process and user calibration (if necessary).

WARNING

Do not use any solutions or standards other than the Hanna 

Instruments

®

 CAL Check Standards. For accurate validation and calibration results, 

please perform these at room temperature, 18 to 25 

°

C (64.5 to 77.0 

°

F).

Note: 

CAL Check Standards will not read the specified value in measurement 

mode. Protect the CAL Check cuvettes from direct sunlight by keeping them in 

the original packing. Store between 5 and 30 

°

C (41 to 86 

°

F), do not freeze.

To perform a CAL Check:

1.  Press 

CAL Check

 from measurement mode. 

The 

Not Available

 message or the date, time and status of the last CAL Check 

will be displayed on the screen.

 

2.  Press 

Check

 to start a new CAL Check. Press the 

 key at any time to abort 

the validation process. 

3.  Use the functional keys to enter the certificate value of the calibration standard 

found on the CAL Check Standard Certificate. Press 

Next

 to continue.

Note: 

This value will be saved in the instrument for future validation. If a new set 

of calibration standards is obtained, please update the certificate value.

4.  Insert the 

A ZERO

 CAL Check Cuvette A then press 

Next

 to continue. The 

Please 

wait...

 message will be displayed during the measurement.

 

Summary of Contents for HI97712

Page 1: ...INSTRUCTION MANUAL HI97712 Aluminum Photometer...

Page 2: ...a Hanna Instruments product Pleasereadthisinstructionmanualcarefullybeforeusingthisinstrument This manual will provide you with the necessary information for correct use of this instrument as well as...

Page 3: ...6 1 METER VALIDATION CAL CHECK CALIBRATION 11 6 2 CHEMICAL FORMULA UNIT CONVERSION 12 6 3 GLP 13 6 4 LOGGING DATA LOG RECALL 13 6 5 GENERAL SETUP 14 6 6 REAGENTS ACCESSORIES 16 6 7 CONTEXTUAL HELP 16...

Page 4: ...Cuvette A HI97712B CAL Check Cuvette B for Aluminum Cloth for wiping cuvettes Scissors 1 5V AA Alkaline batteries CAL Check standard certificate Instrument quality certificate Instruction manual Each...

Page 5: ...urs wipe up immediately and rinse with plenty of water If reagent contacts skin rinse the affected area thoroughly with water Avoid breathing released vapors Waste disposal For proper disposal of reag...

Page 6: ...etector Silicon photocell Cuvette type Round 24 6 mm diameter 22 mm inside Additional Specifications Auto logging 50 readings Display 128 x 64 pixel B W LCD with backlight Auto off After 15 minutes of...

Page 7: ...mance of the instrument at any time and apply a user calibration if necessary Hanna Instruments CAL Check cuvettes are made with NIST traceable standards The built in tutorial mode guides users step b...

Page 8: ...ver Keypad Description The keypad contains 3 direct keys and 3 functional keys with the following functions Press the functional key to perform the function displayed above it on the LCD Press and hol...

Page 9: ...sorbed by atoms molecules or crystal lattices Photometric chemical analysis is based on specific chemical reactions between a sample and reagent to produce a light absorbing compound If pure absorptio...

Page 10: ...luminous efficiency providing more light while using less power They also produce little heat which could otherwise affect electronic stability LEDs are available in a wide array of wavelengths wherea...

Page 11: ...m direct sunlight by keeping them in the original packing Store between 5 and 30 C 41 to 86 F do not freeze To perform a CAL Check 1 Press CAL Check from measurement mode The Not Available message or...

Page 12: ...creen OUT OF SPECIFICATION A user calibration is not allowed the measured value is outside of the tolerance window Check the certified value expiration date and clean the outside of the cuvette Repeat...

Page 13: ...l measurements Every time a measurement is made the data is automatically saved The data log can hold 50 individual measurements When the data log is full 50 data points the meter will rewrite the old...

Page 14: ...ease or decrease the value Press Accept to confirm or the key to return to the Setup menu without saving the new value Contrast Option 0 to 100 Press Modify to change the display s contrast Use the fu...

Page 15: ...ption Comma or Period Press the functional key to select the desired decimal separator The decimal separator is used on the measurement screen Language Press Modify to change the language Use the func...

Page 16: ...ore Factory Settings Press Select to reset to factory settings Press Accept to confirm or Cancel to exit without restoring the factory settings 6 6 REAGENTS ACCESSORIES Press the key to enter the menu...

Page 17: ...o diagnostic test when it is powered on During this test the Hanna Instruments logo will appear on the LCD If the auto diagnostic test was successful the meter is ready for use The battery icon on the...

Page 18: ...hnique is listed in the method procedure The mixing method is indicated with shake gently using one of the following icons In order to avoid reagent leaking and to obtain more accurate measurements cl...

Page 19: ...in the method It is possible to take multiple readings in a row but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement when possible D...

Page 20: ...ee ACCESSORIES section MEASUREMENT PROCEDURE Note If tutorial mode is disabled follow the measurement procedure below If the tutorial mode is enabled press Measure and follow the messages on the scree...

Page 21: ...he cap is positioned securely in the groove Press Zero The display will show a 15 minute countdown prior to zeroing the blank To skip the timer press Zero twice The display will show 0 0 when the mete...

Page 22: ...O3 INTERFERENCES Interference may be caused by Alkalinity above 1000 mg L Phosphate above 50 mg L Iron above 20 mg L Fluoride must be absent If the fluoride concentration is known the aluminum concent...

Page 23: ...mple 4 From the intersection of the fluoride and aluminum line follow the line to the left until it intersects the y axis This point corresponds to the corrected aluminum concentration in the sample E...

Page 24: ...the zero cuvettes are inverted Swap the cuvettes and repeat the measurement There is either too much light or the instrument can not adjust the light level Please check the preparation of the zero cu...

Page 25: ...e meter If the issue persists please contact Hanna Instruments technical support Battery level is too low for the meter to function properly Replace the batteries with new ones Tutorial mode has been...

Page 26: ...batteries follow these steps Turn the instrument off by pressing and holding the key Remove the battery cover by turning it counterclockwise Remove the old batteries replace them with three new 1 5V...

Page 27: ...de Description HI7101412 blue carrying case for HI977xx and 2 CAL Check cuvettes HI731318 cloth for wiping cuvettes 4 pcs HI731331 glass cuvette 4 pcs HI731336N cap for glass cuvette 4 pcs HI93703 50...

Page 28: ...ste Hand them over to the appropriate collection point for recycling Ensuring proper product and battery disposal prevents potential negative consequences fortheenvironmentandhumanhealth Formoreinform...

Page 29: ...equired contact your local Hanna Instruments Office If under warranty report the model number date of purchase serial number engraved on the bottom of the meter and the nature of the problem If the re...

Page 30: ...30 MAN97712 11 19...

Page 31: ...Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice...

Page 32: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN97712 Printed in ROMANIA...

Reviews: