17
Chemical Oxygen Demand Low Range
7. METHOD PROCEDURE
7.1. CHEMICAL OXYGEN DEMAND LOW RANGE
REQUIRED REAGENTS
Code
Description
Quantity
EPA REAGENT
HI93754A‑0
*
COD Low Range EPA Reagent Vial
2 vials
DEIONIZED120
Deionized Water
2 mL
MERCURY FREE REAGENT
HI93754D‑0
*
COD Low Range Hg Free Reagent Vial
2 vials
DEIONIZED120
Deionized Water
2 mL
ISO REAGENT
HI93754F‑0
*
COD Low Range ISO Reagent Vial
2 vials
DEIONIZED120
Deionized Water
2 mL
REAGENT SETS
HI93754A‑25
Reagents EPA Low Range for 25 tests
HI93754D‑25
Reagents Hg Free Low Range for 25 tests
HI93754F‑25
Reagents ISO Low Range for 25 tests
* Reagent vial identification: COD A, COD D, COD F, red label
For other accessories see Accessories section.
Note
: Store the unused vials in their packaging in a cool and dark place.
MEASUREMENT PROCEDURE
Before using the reagent kit carefully read all the instructions and the Safety Data Sheets (SDS).
Pay particular attention to all warnings, cautions and notes. Failure to do so may result in serious
injury to the operator.
Reagent Blank Correction:
This method requires a reagent blank correction. A single blank vial may be used
more than once. The blank vial is stable for several months at room temperature. For improved accuracy, run a
blank for each set of measurements and always use the same lot of reagents for blank and samples.
• Choose a homogeneous sample. Samples containing solids capable of settling need to be homogenized
with a blender.
• Preheat the Hanna
®
Reactor
HI839800
to 150 °C (302 °F). The optional
HI740217
safety shield is
strongly recommended. Do not use an oven or microwave; samples may leak and generate a corrosive
and possibly explosive atmosphere.
• Remove the cap from two COD Low Range Reagent vials.
#1
#2