METHOD PROCEDURE
• Add the content of one packet of
HI93722-0
Cyanuric
Acid Reagent and mix to dissolve.
• Fill the second cuvette with 10 mL of the reacted sample
(up to the mark). Replace the plastic stopper and the
cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Read
. The display will show a 45 second countdown prior to the measurement.
To skip the timer, press
Read
twice. When the timer ends, the meter will perform the
reading. The instrument displays the results in
mg/L
of
cyanuric acid.
10 mL
# 2