•  Handle instrument carefully. Do not drop.
•  Do not immerse instrument under water.
•  Do not spray water to any part of instrument except the “sample well” 

located over the prism.

•  Do not expose instrument or prism to solvents that will damage it. This 

includes most organic solvents and extremely hot or cold solutions.

•  Particulate matter in a sample may scratch the prism. Absorb sample 

with a soft tissue and rinse sample well with deionized or distilled water 
between samples.

•  Use plastic pipettes to transfer all solutions. Do not use metallic tools 

such as needles, spoons or tweezers as these will scratch the prism.

•  To reduce the effects of evaporation or absorption of water when taking 

readings over a period of time, the prism and sample well can be covered 
with plastic wrap.

MEASUREMENT GUIDELINES

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for HI96800

Page 1: ...HI96800 Refractometer for Refractive Index and Brix Measurements INSTRUCTION MANUAL www com information itm com 1 800 561 8187...

Page 2: ...you with the necessary informa tion for correct use of this instrument as well as a precise idea of its versatility Dear Customer All rights are reserved Reproduction in whole or in part is prohibited...

Page 3: ...Description Display Elements Measurement Guidelines Calibration Procedure Measurement Procedure To Change Measurement Unit To Change Temperature Unit Making a Standard Brix Solution Battery Replaceme...

Page 4: ...local Hanna Instruments Office Each HI96800 instrument is supplied with 9 V battery Instruction manual Plastic pipette Quality Certificate Note Save all packing material until you are sure that the i...

Page 5: ...stheuncertainityassociatedwith mechanicalrefractometersandiseasilyportableformeasurementsinthefield The measurement technique and temperature compensation employ metho dologyrecommendedintheICUMSA Met...

Page 6: ...mperature Compensation Automatic between 10 0 and 40 0 C 50 0 104 0 F Measurement Time Approximately 1 5 seconds Minimum Sample Volume 100 L cover prism totally Light Source Yellow LED Sample Cell Sta...

Page 7: ...refractive index there is a critical angle at which an incoming beam of light can no longer refract but will instead be reflected off the interface The critical angle can be used to easily calculate...

Page 8: ...y LCD 2 Secondary Display 3 Primary Display 4 READ Key User Measurement 5 ZERO Key User Calibration 6 RANGE Key User Measurement Unit 7 ON OFF 8 Stainless Steel Sample Well and Prism TOP VIEW BOTTOM F...

Page 9: ...ory Calibration Tag 5 CAL Calibration Tag 6 Measurement Units 7 Automatic Temperature Compensation blinks when temperature exceeds 10 40 C 50 104 F range 8 Temperature Units 9 SecondaryDisplay display...

Page 10: ...rticulate matter in a sample may scratch the prism Absorb sample with a soft tissue and rinse sample well with deionized or distilled water between samples Use plastic pipettes to transfer all solutio...

Page 11: ...ed or deionized water Make sure the prism is completely covered Note If the ZERO sample is subject to intense light such as sunlight or another strong source cover the sample well with your hand or ot...

Page 12: ...ration 3 Press the READ key The result is displayed in the selected units Note The last measurement value will be displayed until the next sample is measured or the instrument is turned off Temperatur...

Page 13: ...y is pressed and the primary display indicated nd for refractive index nd20 for temperature compensated refractive index and bri for Brix when the instrument displays the screen with 4 dashes it is re...

Page 14: ...will display the all segment screen followed by a screen with the model number on the primary display and the version number on the secondary display Continue pressing the ON OFF key 2 While continui...

Page 15: ...r g Total 25 25 000 75 000 100 000 To make a Brix Solution follow the procedure below Place container such as a glass vial or dropper bottle that has a cover on an analytical balance Tare the balance...

Page 16: ...side down and remove the battery cover by turning it counterclockwise Extract the battery from its location Replace with a new 9V battery making sure to observe polarity Insert the back battery cover...

Page 17: ...Cal segment ON Temperature exceeds ATC low limit 10 0 C during calibration t HI primary display Cal segment ON Temperature exceeds ATC high limit 40 0 C during calibration Air Prism surface insufficie...

Page 18: ...18 MAN96800 04 17 ACCESSORIES Code Description HI4020 11 50 Brix Standard 10 mL HI740157P Plastic refilling pipette 20 pcs HI740029P 9V battery 10 pcs www com information itm com 1 800 561 8187...

Page 19: ...the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorizat...

Page 20: ...ters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Local Office Hanna Instruments USA 270 George Washington Highway Smithfield RI 02917 MAN96800 www com in...