19
• Place the
CAL Check
TM
Standard
HI96735‑11
Cuvette A into the cuvette holder and ensure that
the notch on the cap is positioned securely into the groove.
• Press
ZERO/CFM
and the lamp, cuvette and detector icons will appear on the display, depending
on the measurement phase.
• Remove the cuvette.
• Place the
CAL Check
TM
Standard
HI96735‑11
Cuvette B into the cuvette holder and ensure that
the notch on the cap is positioned securely into the groove.
• After a few seconds, the display will show
“-0.0-”
. The meter is now zeroed and ready for
validation.
VALID
ATION PROCEDURE