12
CALIBRATION CHECK ERROR MESSAGE
If during
CALIBRATION CHECK
the meter displays more
than ±0.3 °C (±0.5 °F), then an
“Err”
message is
displayed.
Contact your local Hanna Instruments Office for recalibration.
WARNING MESSAGES
NO PROBE message
If no temperature probe is plugged in, or the signal is
interrupted the display shows blinking dashes and the “
NO
PROBE
” message is displayed.
When the readings are very far from the measurement
range, the probe may be damaged. The display shows
blinking dashes and “
NO PROBE
” message is displayed.
OPERA
TIONAL GUIDE