TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be followed carefully to ensure measurements are conducted
with the highest possible accuracy and precision.
• IT IS CRITICAL THAT THE TIP BE IMMERSED IN THE SOLUTION BEING TITRATED (APPROXIMATELY
0.25 CM).
• Use a clean, volumetric pipette to measure and transfer the necessary volume of sample into
the titration beaker.
• Calibrate the pump prior to each series of titrations.
• Calibrate the pump if the meter is left idle for several hours.
• Analyze the sample immediately after the sample is obtained.
• For better performance, soak the electrode in
HI 70300
storage solution for at least one hour,
before use.
• Press
Restart
to begin a new titration or
ESC
to return to the titration menu.
• If the concentration exceeds the range limits the
exceeded range limit will be displayed blinking.
Another titration can be started by pressing
Restart
.
•
“Wrong input”
error message appears when the
input reading (pH, temperature) exceeds the
specified limits. The pH or temperature value and
concentration will blink indicating an error.
• This screen appears when the stirrer is not working
properly. Check the stir bar and beaker content.
Press
Restart
to try again.
• This error message appears when the pump is not
working properly. Check the tubing, valve and
syringe. Press
Restart
to try again.