Capacitive touchscreen

Users

Settings Log Recall

Help

Channel  

Settings

Back key

Home key

Menu key

•     return to a previous hierarchical menu level

•     access to measurement screen

•     access to:

 

`

Users

 

login & rights configuration

 

`

Settings

  system configuration, connectivity & printing items

 

`

Log Recall

  logged measurement data

 

`

Help

 support

•    from measurement screen, tap to access measurement menu

Probe usage & Calibration tips

Accessories

: rinse container, calibration container, calibration 

standard, DI water

Prior to placing the probe into a standard:

• Remove the transparent protective ring (1).

• Rinse in DI water and gently shake off excess water.

• Immerse the probe in rinse container ensuring that probe’s holes 

(3) are completely submerged. Swirl and remove.

• Immerse and swirl the probe several times to prime the probe.  

3

2

1

Calibration tips

• Ensure the probe is correctly plugged in.

• Use fresh standards with a value close to that of the sample.

• Pour fresh standard in calibration container, immerse the probe, 

gently swirl, and remove.

• Immerse and swirl the probe several times.

• Gently tap the probe to remove trapped air bubbles.

• Wait for the temperature to stabilize and the probe to acclimate 

to the solution.

• Calibrate frequently and discard standard after use.

Conductivity measurement tips

• Work with recently calibrated probes.

• Substitute calibration standard with measured sample and 

follow calibration recommendations.

• Use the arm holder for easy probe transfer in and out of 

container and correct probe positioning, away from container 

walls or bottom.

• TDS measurement: use correct TDS conversion factor for 

measured sample.

• Record measured data once the reading is stable.

• Clean and rinse off any residual sample on the sensors (2) after 

measurement to keep the probe in optimum condition.

Logging in & Creating a new account 

1. Power the meter and tap

 

.

2. Tap 

  followed by

 

 .

3. Tap 

Edit Users

 to enter Account Management.

4.  Tap   to enable Account Creation and Logins. Tap   to return.

5. Tap Logout to enter Users screen. Tap the plus symbol avatar.

6.  Input user name and tap

 

.

7. Enter password and tap

 

. Re-enter password to confirm.