background image

Instruction Manual

Manufactors  since 1978

These Instruments are in

Compliance with the CE

Directives

HI 150 Series

Thermometers

Dear Customer,
Thank you for choosing a Hanna product. This manual will

provide you with the necessary information for correct

operation. Please read it carefully before using the meter. If

you need additional technical information, do not hesitate

to e-mail us at [email protected].
These instruments are in compliance with the directives EN

50081-1 and EN 50082-1.

Remove the instrument from the packing material
and examine it carefully. If any damage has oc-
curred during shipment, immediately notify your
Dealer or the nearest Hanna Customer Service Cen-

ter.
The meters are supplied with 12 VDC power adapter.

Note:

Conserve all packing material until the in-

strument has been observed to function
correctly because any defective item must be
returned in its original packing.

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

GENERAL DESCRIPTION

GENERAL DESCRIPTION

GENERAL DESCRIPTION

GENERAL DESCRIPTION

GENERAL DESCRIPTION

HI 150 is an indicator with a relay Output and
built-in LED Alarm Signal for the monitoring of
temperature.
Constructed in a compact casing and with a molded
eye, HI 150 can be installed on the QC bench or
wherever necessary to control the ambient tem-

perature of food chain, from the storage room to
the point of delivery.

The housing has been completely sealed against

vapors and humidity.

The measurement range is from -20 to 60ºC  with
an excellent ±0.5ºC  accuracy. The stainless steel
encapsulated sensor is positioned right on the
outer body of this thermometer to provide a rapid
response.
In order to ensure that temperature control is
never interrupted, HI 150 runs permanently on a
12 VDC power supply. Temperature is constantly
displayed on a large LCD so that the operator  can

check it at any time.

HI 150 also offers  a unique alarm control. Any
value from 0  to 50.0ºC (HI 150-00) or  -20 to
20ºC (HI 150-10) can be adjusted as the setpoint
through a trimmer located on the back of the
meter. Should the temperature vary from this user-
selectable setpoint, the red LED blinks and the
relay is activated to warn of an abnormality or
switch on a  system. The built-in LED is well visible

from a distance.
This alarm relay is a cost-saving feature with a
tremendous potential that can help you in auto-
mating your  temperature control. It can also be
used as a forewarning that the refrigeration  or air
conditioning  unit is malfunctioning, or that a door
has been left open, or point to any other mishaps,
common in a busy workplace.
HI 150 is a convenient and  inexpensive  way to
ensure  that temperature  never varies too much
from acceptable levels. It frees workers  from hav-
ing to manually check the temperature and
immediately signals if and when an abnormality
arises.

S P E C I F I C A T I O N S

S P E C I F I C A T I O N S

S P E C I F I C A T I O N S

S P E C I F I C A T I O N S

S P E C I F I C A T I O N S

HI150-00

HI150-00

HI150-00

HI150-00

HI150-00

Range

-20.0 to 60.0ºC

Resolution

0.1ºC

Accuracy

±0.5ºC

Setpoint

Selectable from 0.0 to 50.0 ºC

Hysteresis

±0.5ºC around setpoint

Alarm

          LED blinks and alarm relay activated

                when measured temperature is
                lower than the setpoint

Relay          Through contacts inside the meter
Sensor

        Incorporated, encapsulated stainless

               steel

Environment 

          -20 to 60ºC; RH 100%

Power supply

External 12 VDC (included)

Dimensions

86 x 94 x 33mm (3.4 x 3.7 x 1.3”)

Weight

150g (5.2oz.)

HI150-10

HI150-10

HI150-10

HI150-10

HI150-10

Range

-20.0  to  60.0ºC

Resolution

0.1ºC

Accuracy

±0.5ºC

Setpoint

Selectable from -20.0  to 20.0 ºC

Hysteresis

±0.5ºC  around setpoint

Alarm

 LED blinks and alarm relay activated

                when measured temperature is
                higher than the setpoint

Relay          Through contacts inside the meter
Sensor

        Incorporated, encapsulated stainless

               steel

Environment 

       -20 to 60ºC; RH 100%

Power supply

External 12 VDC (included)

Dimensions

86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)

Weight

150g (5.2oz.)

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

TAKING MEASUREMENTS

• Turn the meter on by connecting the 12 VDC

power adapter to the meter and to the mains.

• A value will appear on the LCD  to indicate the

temperature. Wait for a few minutes  for the
display to stabilize.

ADJUSTING THE SETPOINT

With the 

HI 150

, you can select your own setpoint

and be alerted with a visual LED alarm when an
abnormal situation arises.
• Unscrew and remove the rear panel and gasket

seal to access the MEASURE/SET switch.

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

1. Molded eye
2. Alarm LED
3. Access for relay contacts
4. 12VDC power adapter
     (included)
5. Power supply connector
6.  Sensor
7. Liquid Cristal Display

Reviews: