HI10482/HALO pH PROBE  

QUICK START INSTRUCTION GUIDE 

 

Thank you for selecting a Hanna Instruments product. Please read this manual before using the Halo     
probe on your iPad® with Hanna Lab App or with edge® blu.  Consult the Hanna Lab App "HELP" or the 
edge blue instruction manual for additional information on  Probe Maintenance and calibration. 

IMPORTANT!

  During servicing and measurement,

 

handle probe with care. Do not drop probe on a hard 

surface.  It will permanently damage the probe.

 During storage or transport, utilize the clear protective cap 

to protect and keep the pH bulb hydrated. 

PREPARATION:  

Remove the probe from the plastic storage tube by unscrewing 
the top collar to vent the cap before removing.  Save cap and 
collar for probe storage. 
1.

 

Rinse off any storage solution or salts that may be on the 
glass body.  

2.

 

Remove the fill hole cap and top off electrolyte reservoir 
with HI 7082 if required. It is best to operate the probe 
with fill hole open to ensure electrolyte flow through the 
junction.  

3.

 

Check to verify there is solution inside the pH bulb, 
shaking the probe down can restore continuity as the 
solution may have moved up the stem during shipping. 

 
CONNECTION

Using Hanna Lab App on an iPad® 
1.

 

To begin using the HALO™ probe, start the Hanna Lab 
app and tap the Bluetooth icon on the iPad®.   

2.

 

Press the operation button located on the top of the probe. 
The blue LED halo will begin to flash in half second 
intervals, indicating it's in discoverable mode and the 
probe ID will appear in the list of probes. 

3.

 

Touch the probe ID once it appears in the list of probes. 
The blue LED will slow down and flash every 2 seconds. 

 
 
Using Edge®blu 
1.

 

To begin using the HALO™ probe, turn ON edge®blu. 
The meter swill start scanning for probes. 

2.

 

Press the operation button located on the top of the probe. The blue LED halo begin to flash in half 
second intervals, indicating it's in discoverable mode and the probe ID will appear in the list of probes. 

3.

 

Select your probe in the list of visible probes and press CFM. The blue LED will start flashing 4 times 
a second while connecting, then it will slow down and flash every 2 seconds showing it is connected 
and ready to measure. 

 
 

MEASUREMENT SETUP

1.

 

Calibrate the pH probe. 

2.

 

Take measurements using the pH probe. 

Hole

Domed pH Sensitive Bulb

Movable PTFE Sleeve,
Open Junction

Operation Button

Blue HALO Indicator