Hanna Instruments HI1006-18 Instruction Manual Download Page 10

Wiring the Probe to the Controller 

10

9.  WIRING THE PROBE TO THE CONTROLLER

1.  With the controller disconnected from power, run the probe cable through the conduit opening.
2.  Connect the probe leads to the removable terminal connector marked PROBE. Follow the lead markings 

(+) positive / (–) negative to ensure correct wiring position for the output leads.

3.  Carefully put the wired terminal connector into place on the board.
4.  Position the excess cable through the cable gland before tightening the nut.
5.  Remove the ground screw and hardware located below the PROBE connector and attach ground lead ( 

).

COMM

- B A +

IN1 IN2  A01 A02  A03 A04 

PROBE

- B A +

PROBE

- B A +

A +

B

Probe cabling color code

Marking

Attached cable

Patch cable *

Functionality

GREEN

BLACK

0V

B

WHITE

WHITE

RS485 D –

A

YELLOW

BLUE

RS485 D+

+

BROWN

RED

5V

g

GREEN-YELLOW 

Protective ground connection

See Accessories section for cable ordering codes.

Note

: Ensure wiring regulations are correctly followed when controller unit is part of a larger industrial installation.

10.  CALIBRATION

HI510

 Controller allows two types of pH calibration procedures:

• 

Standard

 calibration performed in standard buffer solutions

Hanna Instruments

®

: 1.68, 4.01, 7.01, 10.01, 12.45 pH

NIST: 1.68, 4.01, 6.86, 9.18, 12.45 pH

• 

Process

 calibration performed with probes calibrated in standards and installed in the process

Summary of Contents for HI1006-18

Page 1: ...rature Industrial Smart Probes SN XXXXXXXXXXXX pH 0 12 T 5 to 80 C p max 6 bar HI1006 1805 SN XXXXXXXXXXXX pH 0 12 T 5 to 80 C p max 6 bar HI1006 1800 Please scan the QR code or follow the link to dow...

Page 2: ...ssories 13 13 Abbreviations 14 Certification 14 Recommendations for Users 14 Warranty 14 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully...

Page 3: ...trial pH probes intended for industrial process control when teamed up with the HI510 Universal Process Controller HI1006 18 and HI1016 18 series designed for low conductivity or low temperature proce...

Page 4: ...uoride resistant HF glass sensor titanium matching pin 5 0 to 60 0 C 23 0 to 140 0 F 0 00 to 10 00 pH 8 Smart probe with RS485 connection zz 00 supplied with DIN connector without cable See Accessorie...

Page 5: ...controller setting Temperature compensation Automatic 0 0 to 80 0 C 32 0 to 176 0 F Temperature source Automatic from Probe Manual Body PVDF Junction PTFE Sensor Body LT glass Tip shape flat self cle...

Page 6: ...ations Probes are easily installed using the 3 4 NPT external thread Hand tighten the probe in position Then depending on the process tighten one or two turns with a wrench to secure in place Do not e...

Page 7: ...quire that saddle and flow cell are completely filled with water In line mounting with probe saddle 1 2 4 3 5 Note Probe can be rotated to prevent air entrapment inside the electrode Mounting accessor...

Page 8: ...enance and replacement B Slide out the electrode holder cap 3 C Screw the probe into the internally threaded protective cap 9 step I D Screw the probe and protective cap subassembly onto the holder 5...

Page 9: ...sediment or other foreign material to accumulate within the sensing area TANK VALVE 1 VALVE 2 FLOW FLOW INCOMING FLOW EXIT FLOW 3 1 2 4 5 7 6 6 FLOW CELL 1 Process controller 2 Wiring cable 3 pH prob...

Page 10: ...tor and attach ground lead g COMM B A IN1 IN2 A01 A02 A03 A04 PROBE B A PROBE B A A B Probe cabling color code Marking Attached cable Patch cable Functionality GREEN BLACK 0V B WHITE WHITE RS485 D A Y...

Page 11: ...CFM to save Wait followed by Calibration point is accepted and saved is displayed after the calibration point has been saved and accepted 5 Select Next to continue with a two point calibration 6 Repea...

Page 12: ...ubbles inside the glass bulb If the bulb and or junction are dry soak the electrode in HI70300 Storage solution for at least 30 minutes To ensure a quick response time the glass bulb and the junction...

Page 13: ...L Oil and fat cleaning solution 230 or 500 mL Other Accessories HI60501 PVC immersion electrode holder HI60503 PVDF immersion electrode holder HI60501 0 O rings for HI60501 electrode holder HI605011 P...

Page 14: ...roduced by the user to the supplied equipment may degrade its performance Do not use or store the product in hazardous environments WARRANTY The probes are warrantied for six months against defects in...

Reviews: