ELECTRODE CONDITION

The display is provided with a probe icon (unless the feature 
is disabled from setup) which indicates the electrode status 
after calibration. The “condition” remains active for 12 
hours (unless the batteries are removed).
The electrode condition is evaluated only if the current 
calibration has two points.
 

5 bars: excellent condition

 

4 bars: very good condition

 

3 bars: good condition

 

2 bars: fair condition

 

1 bar: poor condition

 

1 bar blinking: very poor condition

With 1 bar it is recommended to clean the electrode and 
recalibrate. If there is still only 1 bar or 1 bar blinking 
replace the probe.

Sensor check

Setting the meter to pH‑mV range the user can check the 
sensor status at any time. The offset value is the reading 
in pH 7.01 buffer (@ 25 °C/77 °F). If this reading is 
outside the range ±30 mV, the electrode is considered 
“very poor”. The slope value of the sensor is the difference 
between readings in pH 7.01 and in pH 4.01 buffers. When 
the slope reaches the value of about 150 mV, the electrode 
is considered “very poor“. When “poor“ or “very poor“, it 
is recommended to replace it with a new one.

Note: To ensure reliable readings, the electrode must 

be cleaned with cleaning solution and then hydrated in 

storage solution for a minimum of 30 minutes before 

calibrating the probe.

OPERA

TIONAL GUIDE

Summary of Contents for HI 99163

Page 1: ...HI99163 Waterproof pH Temperature Meter for Meat INSTRUCTION MANUAL...

Page 2: ...a Hanna Instruments product Please read this instruction manual carefully before using this meter This manual will provide you with the necessary information for correct use of this meter as well as...

Page 3: ...ral description and Intended use Main features Specifications Display description Operational guide Meter setup Battery replacement Accessories Electrode maintenance Certification Recommendations for...

Page 4: ...ith FC2323 pH temperature probe with stainless steel blade DIN connector and 1 m 3 3 cable HI70004 pH 4 01 buffer 1 sachet HI70007 pH 7 01 buffer 1 sachet HI700630 grease and fats acid cleaning soluti...

Page 5: ...a large dual line display and automatic pH calibration at one or two points The FC2323 probe has a removable stainless steel blade that allows pH measurements to be taken in intact muscle during proce...

Page 6: ...wo memorized buffer sets standard or NIST Selectable temperature unit C or F Electrode condition indicator mV of pH measurement for electrode check FC2323 dedicated pH with integrated temperature sens...

Page 7: ...set Standard 4 01 7 01 10 01 or NIST 4 01 6 86 9 18 Probe included FC2323 preamplified pH and temperature probe with stainless steel blade DIN connector and 1 m 3 3 cable Battery type life 1 5V AAA 3...

Page 8: ...mperature Compensation indicator 2 Stability indicator 3 Battery percentage 4 Electrode condition indicator 5 Low battery indicator 6 Measurement unit 7 Primary LCD 8 Temperature unit 9 Secondary LCD...

Page 9: ...lled in their place The meter is provided with an active acoustic signal when a key is pressed At start up the meter displays all LCD segments for a few seconds followedbythepercentageindicationofther...

Page 10: ...LECTING MEASUREMENT RANGE While in measurement mode press the SET button to select pH or pH mV measurement on the first LCD line FREEZING MEASUREMENT VALUES While in measurement mode press and hold th...

Page 11: ...the first calibration buffer If performing a two point calibration use pH 7 01 pH 6 86 for NIST buffer first The meter will enter the calibration mode displaying pH 7 01 USE and tag blinking or pH 6 8...

Page 12: ...calibration For better accuracy it is always recommended to carry out a two point calibration After the calibration procedure has been completed the tag is turned on together with the calibrated poin...

Page 13: ...re is still only 1 bar or 1 bar blinking replace the probe Sensor check Setting the meter to pH mV range the user can check the sensor status at any time The offset value is the reading in pH 7 01 buf...

Page 14: ...s to confirm and to enter the BEEP selection To switch the beep tone ON or OFF press the SET button press to confirm and to enter the calibration buffer selection pH 7 01 BUFF To change the buffer set...

Page 15: ...display shows bAtt DEAD for a few seconds then the meter powers off Immediately replace the batteries with new ones The batteries are accessed by opening the battery cover on the back of the instrumen...

Page 16: ...pH 10 01 buffer solution 500 mL HI70300L pH electrode storage solution 500 mL HI7061L pH electrode cleaning solution 500 mL HI70630L Grease and fats acid cleaning solution 500 mL HI70631L Grease and f...

Page 17: ...orage Solution Follow PREPARATION above before taking measurements Note Never store the electrode in distilled water PERIODIC MAINTENANCE Inspect the electrode for any scratches or cracks If any prese...

Page 18: ...hich will conserve natural resources Disposal of waste batteries This product contains batteries do not dispose of them with other household waste Hand them over to the appropriate collection point fo...

Page 19: ...d by the warranty you will be notified of the charges incurred If the meter is to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Technical Service departm...

Page 20: ...Printed in ROMANIA World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN99163...

Reviews: