11
METER SETUP
METER SETUP
Setup mode allows the selection of the Temperature unit, Auto‑off, Beep confirmation tone, the type of
pH buffer set, the Resolution and Information. To enter Setup mode press and hold button until
“
STD
“ and
tag is replaced by “
SETUP
“ and
tag. Release the button.
• “
TEMP
“ is displayed on the secondary LCD with the current temperature unit
(e.g. “
TEMP °C
“), for °C/°F selection, use the
SET
button. After the temperature
unit has been selected, press to confirm and to enter the “
A-OFF
“ selection.
• Use the
SET
button, to navigate through the auto‑off choices: 8 minutes
(
“8”
, default value), 60 minutes (
“60”
) or disabled (
“---”
). Press to
confirm and to enter the “
BEEP
“ selection.
• To switch the beep tone ON or OFF, press the
SET
button; press to confirm
and to enter the calibration buffer selection “
pH 7.01 BUFF
“.
• To change the buffer set, the meter will show the current buffer set: “
pH 7.01
BUFF
” (for standard buffer set: 4.01/7.01/10.01) or “
pH 6.86 BUFF
” (for
NIST buffer set: 4.01/6.86/9.18). Change the set with the
SET
button. Press
to confirm and to enter pH resolution selection “
RESOL
“.
• To change the pH measurement resolution between “
0.1
“ and “
0.01
“ use
the
SET
button; then press
to confirm and to enter electrode calibration
information “
INFO
“ selection.
• To switch the electrode condition indicator on the LCD ON or OFF, press the
SET
button; press to exit setup options. Change the set with the
SET
button,
then press
to confirm and to return to normal mode.