When D.O. percentage saturation is calibrated, also D.O. concentra-
tion is calibrated as a result and viceversa.
6.4 CONDUCTIVITY
• From the GLP data main menu, select Con-
ductivity entry and press OK. In this menu it is
possible to visualize calibration data for conduc-
tivity, actual conductivity and salinity.
• All the information relative to the last Conductivity calibration appear: the
calibrated point, the cell constant value, the calibration type (conductivity, actual
conductivity or salinity), time and date.
• With the arrow keys it is possible to scroll between the last 5 memorized cali-
brations.
For Conductivity GLP calibration data different screen types are possible: con-
ductivity, actual conductivity and salinity.
NOTES:
The C label near the calibration point is for custom and the H is for
Hanna.
6.5 ATMOSPHERIC PRESSURE
• From the GLP data main menu, select the Atmospheric pressure entry and
press OK.
• All the information relative to the last Atmo-
spheric pressure calibration appear: custom cali-
bration point, time and date; the message “Fac-
tory calibr. restored” in case of restoring factory
calibration data.
• With the arrow keys it is possible to scroll between the last 5 memorized cali-
brations.