background image

11

10

code 40: Auto OFF
code 41: Battery test
code 50: Baud rate
code 60:  Release code
code 70:  Celsius or Fahrenheit
code 99:  Password

Some of the most important functions are explained below in a step
by step sequence.

TO SCAN LOGGED DATA

CODE 00 - Lot Data Printing / Scrolling

• Select the code 00.
• The message "Lot data Printing" will scroll twice on LCD.

• The upper LCD will then display L 00

with the 00 blinking.

• Set the desired lot by the arrow keys. L00 is the lot of data of the

"log on demand" and L01 to L16 are the lots of the "timed log".

• Press the ALT and CFM keys to confirm the lot number.
• If the lot doesn't contain data, the "no data" message will scroll

across the LCD twice and the meter will return to setup mode.

• If the lot contains one or more data the

LCD will display the sample number in its
upper part and Sn in the lower part.

Note:

In the L00 lot (log on demand) the sample number will be
displayed with 3 digits (001).

• Select the sample number to scan by the arrow keys.

Printing Logged Data

• Press ALT and CFM to print logged data.
• If the selected sample number is invalid (equal to 0 or bigger

than the number of samples), the "Err" message will be displayed
for a few seconds.

• If the sample number is correct, the samples starting from the

selected one to the last sample of the lot will be printed. To stop
printing before the last sample is reached, press and hold the ALT
and PAPER keys until the printer stops.

• During printout the LCD will display the sample number being

Code

Valid values

Default

10 Show pH GLP data
11 Show ORP GLP data (

HI 98240

 only)

12 Print GLP data
13 Calibration alarm time-out

00 to 99 days

00

00 means option disabled

14 Instrument ID code

0000 to 9999

0000

30 Current time

1

hh:mm

00:00

31 Current day

1

dd

01

32 Current month

1

MM

01

33 Current year

1

YYYY

1998

40 Auto-Off/Power down time-out 5,10,15,30,45,60

5

41 Battery level test
50 RS232 baud rate

1200, 2400, 4800

4800

60 Firmware version
70 Celsius/Fahrenheit selection

°C or °F

°C

99 Password

2

0000 to 9999

0000

Note:

If a wrong code is entered the "Err" message is displayed on
LCD for a few seconds.

1

The meter automatically checks for entered time/date accuracy as follows:

      0

hh

23; 0

mm

59; 01

dd

28/29/30/31; 1

MM

12; 1998

YYYY

2097.

2

To change the password, the correct code must be entered first. If the password
has been forgotten, the password protected features are no longer accessible;
in this case contact your nearest Hanna Service Center.

SETUP MESSAGES LIST

code. 00: Lot data Printing
code 01: Lot table Printing
code 02: Printer enable
code 03: Log Interval
code 05: Press "ALT CFM" to delete Lot 00 or "ALT FNC" to escape
code 06:  Press "ALT CFM" to delete Lot  01-16 or "ALT FNC" to escape
code 10: pH GLP
code 11: Volt GLP (

HI 98240

 only)

code 12: GLP data printing
code 13: Calibration alarm time-out
code 14: Instrument ID Code
code 30: Hour - Minute
code 31: Day
code 32: Month
code 33: Year

Summary of Contents for HI 98230

Page 1: ...Instruction Manual HI 98230 HI 98240 Printing and Logging pH Meters with GLP and Smart Electrodes w w w h a n n a i n s t c o m...

Page 2: ...meters are also equipped with a stability indicator and backlight feature for comfortable reading even in dark environments The GLP features provide a guarantee of data consistency Measurements can b...

Page 3: ...the LOBAT indication is displayed to warn the user of this condition However the meter continues to measure correctly even when the low battery indication is displayed When the batteries become too we...

Page 4: ...lectrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach the electrode from the meter before immersing the electrode in the storage solution ON OFF SPEC...

Page 5: ...iption of the specific setup items their valid values and the factory settings default Code Valid values Default 00 Lot data printing scrolling 00 to 16 00 01 Print lots data summary 02 Printer enable...

Page 6: ...sample is reached press and hold the ALT and PAPER keys until the printer stops During printout the LCD will display the sample number being Code Valid values Default 10 Show pH GLP data 11 Show ORP G...

Page 7: ...age GLP data printing will scroll across LCD The meter will then print a complete set of the following GLP data a Instrument identification ID code printed at that moment If printout is stopped the LC...

Page 8: ...ferent samples it is recommended to rinse the electrode thoroughly with deionized water or if not available tap water first and then with some of the next sample to condition the electrode before imme...

Page 9: ...ORP electrode to the meter and press ON OFF to power on the instrument The meter automatically sets the mV range If the ORP electrode is not connected the LCD will display the no probe message and th...

Page 10: ...ng will be printed beside the date Note A dashed line indicates that the calibration has never been performed 6 If password is correct or not set the meter will display 7 01pH on the up per LCD and b1...

Page 11: ...when LCD is displaying mV 3 User is prompted to enter password if it has been set to a value different from 0000 otherwise skip to step 6 4 Enter the password with the arrow keys 5 Press ALT and CFM t...

Page 12: ...ly enabled in order to return to logging with printing mode LOG ON DEMAND In measuring or Timed log mode press LOG to store the current reading The LCD will display Two different modes to print log da...

Page 13: ...the meter on again CALIBRATION ALARM TIME OUT The calibration alarm time out is available only for pH calibration It is possible to set through setup code 13 the number of days before the next requir...

Page 14: ...t is possible to make any elaboration available with software e g graphics statistic analysis HI 92000 offers a variety of features and is provided with an on line help to support the user throughout...

Page 15: ...lot was the 16th the new logging period restarts from lot 01 overwriting previously logged data When Timed logging memory is full the meter overwrites the oldest lot data progressively reducing the o...

Page 16: ...I 98240 can also be used with other adapters In this case remember to check the correct polarity of your adapter before connecting it to the meter BATTERY REPLACEMENT TEMPERATURE RESISTANCE CORRELATIO...

Page 17: ...own could be due to Clogged Dirty Junction refer to the Cleaning Procedure above Loss of shielding due to low electrolyte level in refillable electrodes only refill with fresh HI 7071 for single junct...

Page 18: ...7 01 buffer solution 500 mL FDA bottle HI 8009L pH 9 18 buffer solution 500 mL FDA bottle HI 7009L pH 9 18 buffer solution 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle HI 8010L pH 10...

Page 19: ...10005 Voltage adapter 115 Vac 12 Vdc HI 710006 Voltage adapter 230 Vac 12 Vdc HI 710031 Rugged carrying case HI 710034 Paper roll 10 pcs HI 710035 Ink cartridge HI 76405 Electrode holder HI 8427 pH mV...

Page 20: ...s in workmanship and materials when used for their intended purpose and maintained according to instructions The electrodes and the probes are warranted for a period of six months This warranty is lim...

Page 21: ...42 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 0...

Reviews: