53
•
When the reading is stable and close to the selected calibration point, the
CFM
functional key is displayed.
•
Press
CFM
to confirm. The instrument returns to the main screen.
Note:
•If the reading is not close to the selected calibration point or the
difference between first selected point and second selected point is
less than 25 ºC,
“Wrong”
message will blink.
• If the
WRONG
source is the difference between calibration points
increase the temperature of the vessel with hot water in order to be
acceptable.
• If the
WRONG
source is the temperature reading change the probe
and restart calibration.
If calibration cannot be performed contact Hanna Service.
• If a temperature calibration has been previously performed it is
possible to erase the calibration by pressing the
Clear
functional
key.
•
The
“Calibration cleared”
message will be displayed for a few
seconds and the meter will return to the main screen.
•For one point calibration press
ESC
after first point was confirmed.
The instrument will return to the main screen and will memorize
the calibration data.