33
Setup mode allows viewing and modifying the measurement parameters.
The following table lists the
SETUP
parameters, their valid range and the
factory default settings.
Item
Description
Valid Value
Default
Calibr. Time-out Number of days after
Disabled, 1 to 7 days
Disabled
calibr. warning is displayed
Salinity
The solution’s salt content 0 to 70 g/L
0 g/L
BOD
Configuration
-Sample min The minimum diff. between 0.00 to 50.00 mg/L
0.00 mg/L
delta DO
the start and the end DO val.
-Sample min The minimum end DO value 0.00 to 50.00 mg/L
0.00 mg/L
end DO
-Seed min The minimum diff. between 0.00 to 50.00 mg/L
0.00 mg/L
delta DO
the start and the end DO val.
-Seed min The minimum end DO value 0.00 to 50.00 mg/L
0.00 mg/L
end DO
OUR
Configuration
-Min time
The minimum time for the 1 to 3600 sec.
1 s
OUR test
-Max time
The maximum time for the 1 to 3600 sec.
3600 sec.
OUR test
-Min start DO The minimum DO value 0.01 to 50.00 mg/L
0.01 mg/L
for starting the OUR test
-Min end DO The minimum DO value at 0.00 to 50.00 mg/L
0.00 mg/L
the end of the OUR test
-Total volume The total volume of the 0.1 to 300.0 mL
0.1 mL
solution to be tested
-Sample volume The volume of sample in 0.1 to 300.0 mL
0.1 mL
the solution to be tested
SOUR
configuration
-Min time
The minimum
1 to 3600 sec.
1 sec.
time for the SOUR test
-Max time
The maximum
time for the SOUR test
1 to 3600 sec.
3600 sec.
-Min start DO The minimum DO value
for starting the SOUR test 0.01 to 50.00 mg/L
0.01 mg/L
-Min end DO The minimum DO value at 0.00 to 50.00 mg/L
0.00 mg/L
the end of the test
SETUP
SETUP
SETUP
SETUP
SETUP