3 5
• the calibration time,
yymmddhhmmss
(12 chars).
• ISE calibration data (if available), which contains:
• the number of calibrated standards (1 char)
• the ion charge, with sign (2 chars)
• the calibration slope, with sign and decimal point (7 chars)
• the calibration time,
yymmddhhmmss
(12 chars)
• standards information (for each standard)
• type (1 char): 0 - always standard solution.
• status (1 char):N (new) - calibrated in last calibration;
O (old) - from an old calibration.
• warnings during calibration (2 chars): 00 - no warning.
• standard value, with sign and decimal point (7 chars).
• calibration time,
yymmddhhmmss
(12 chars).
PAR
Requests the setup parameters setting.
The answer string contains:
• Instrument ID (4 chars)
• Calibration alarm time out (2 chars)
• SETUP information (2 chars): 8 bit hexadecimal encoding.
• 0x01 - beep ON (else OFF)
• 0x04 - degrees Celsius (else degrees Fahrenheit)
• 0x08 - Offset calibration (else Point calibration)
• Auto-off/Light time (3 chars)
• Auto power off time (3 chars)
• The number of custom buffers (1 char)
• The custom buffer values, with sign and decimal point,
for each defined custom buffer (7 chars)
• The ion charge (2 chars)
NSLx
Requests the number of logged samples (4 chars).
The command parameter (1 char):
• P - request for pH range
• M - request for mV and Rel mV ranges
• I - request for ISE range
LODPxxx
Requests the xxx
th
pH record logged data.
LODMxxx
Requests the xxx
th
mV/Rel mV record logged data.
LODIxxx
Requests the xxx
th
ISE record logged data (
HI 98172
).
LODPALL
Requests all pH Log on demand.
LODMALL
Requests all mV/Rel mV Log on demand.