ACCESSORIES

HI 1286

Double junction, plastic body pH electrode with 2
m (6.6’) cable and BNC connector

HI 7634*

TDS probe with 2 m (6.6’) cable

HI 70004P

pH 4.01 solution, 20 mL sachet (25 pcs)

HI 7004M

pH 4.01 solution, 230 mL bottle

HI 70007P

pH 7.01 solution, 20 mL sachet (25 pcs)

HI 7007M

pH 7.01 solution, 230 mL bottle

HI 70010P

pH 10.01 solution, 20 mL sachet (25 pcs)

HI 7010M

pH 10.01 solution, 230 mL bottle

HI 70032P

1382 ppm solution, 20 mL sachet (25 pcs)

HI 7032M

1382 ppm solution, 230 mL bottle

HI 70442P

1500 ppm solution, 20 mL sachet (25 pcs)

HI 70442M

1500 ppm solution, 230 mL bottle

HI 70300M

Storage solution, 230 mL bottle

HI 700661P

Cleaning solution, 20 mL sachet (25 pcs)

HI 7061M

Cleaning solution, 230 mL bottle

HI 710005

115 Vac / 12 Vdc power adapter

HI 710006

230 Vac / 12 Vdc power adapter

HI 731326

Calibration screwdriver (20 pcs)

To be replaced by authorized technical personnel only

TAKING MEASUREMENTS

• Turn the meter on by connecting the supplied 12 Vdc

power adapter to the meter and then to the mains.

• Remove the protective cap

from the pH electrode.

• Immerse the tips of pH elec-

trode (at least 4cm/1½”),
TDS and temperature probes
into the solution to be tested.
For best accuracy, the probes
should not touch or stand
close to the vessel’s walls or
bottom.

• The three backlit LCDs will show

the pH, TDS and tempera-
ture values, respectively.
Any initial variation in readings may be due to the pH
electrode conditioning and to the temperature compensa-
tion of TDS measurements. Allow the readings to stabilize.

pH ELECTRODE MAINTENANCE

• When not in use, rinse the electrode with water and store it

with a few drops of 

HI 70300

 storage solution in the

protective cap. Always replace the protective cap after use.

DO NOT USE DISTILLED OR DEIONIZED WATER FOR STOR-
AGE PURPOSES.

• If the electrode has been left dry, soak the tip in 

HI 70300

storage solution overnight to reactivate it.

• To minimize clogging and provide longer life for the pH

electrode, it is recommended to clean it at least once a
week. Immerse the tip of the electrode in the 

HI 700661

cleaning solution supplied with the instrument, for half an
hour and then rinse it with tap water.

TDS PROBE MAINTENANCE

It is recommended to clean the probe at least once a month,
by following this procedure:
• Immerse the tip of the probe in the 

HI 700661

 cleaning

solution supplied with the instrument, for half an hour.

• If a more thorough cleaning is required, brush the metal

pins with very fine sandpaper.

• After cleaning, rinse the probe with tap water and recalibrate

the instrument.

pH CALIBRATION

Pour small quantities of pH 7 (

HI 7007

) and pH 4 (

HI 7004

)

or pH 10 (

HI 7010

) buffer solutions into two clean beakers.

For accurate calibration use two beakers for each buffer solu-
tion, the first one for rinsing the electrode tip, and the second
one for calibration.
• Turn the meter on.
• Remove the electrode protective cap, rinse and immerse the

electrode in the pH 7 buffer solution together with the TDS
probe (working as matching pin). Stir gently and wait for
the reading to stabilize.

Note:

 The electrode should be submerged approximately 4 cm

(1½”) in the solution.

• Adjust the “pH 7” trimmer with the supplied calibration

screwdriver until the LCD shows pH “7.0”.

Rinse and immerse the pH electrode and the TDS probe in
the pH 4 (or pH 10) buffer. Stir gently.

• Wait for a stable reading and then adjust the “pH 4”

trimmer to read pH “4.0” (or pH “10.0”).

The pH calibration is now complete.

TDS CALIBRATION

• Turn the meter on.
• Pour a small quantity of the proper calibration solution (HI

7032

 for 

HI 981504/5

 or 

HI 70442

 for 

HI 981504/7

) in a

beaker. If possible, use plastic beakers to minimize any EMC
interference.

• Immerse the TDS probe in the solution, making sure that

the metal pins are completely submerged.

Note:

 For best accuracy, the probe body should not touch nor

stand close to the side walls of the beaker.

• Wait for a couple of minutes for thermal equilibrium to be

reached and the temperature to be compensated.

• Tap the probe gently on the bottom and shake it, to make

sure no air bubbles have remained trapped inside.

• Adjust the calibration trimmer with the supplied screwdriver

until the display shows “1380” ppm (

HI 981504/5

) or

“1500” ppm (

HI 981504/7

)

Calibration is now complete.

CE DECLARATION OF CONFORMITY

ISTCOMBO3-R1     05/05

pH ELECTRODE CONNECTION

In order to protect the instrument against vapors
and humidity, the BNC connector is protected with
a waterproof sheath.
• Slide the protective sheath down. Connect the

pH electrode to the BNC socket and then slide
the protective sheath back up. For maximum
waterproof protection, make sure the connector
is completely covered.

• Do not be alarmed if white crystals appear around the

electrode protective cap. This is normal with pH electrodes
and they dissolve when rinsed with water.

°C/°F SELECTION

• The user can easily select the temperature unit, in degrees

Celsius (default) or Fahrenheit through a switch on the rear
panel.

• Remove the label and set the switch to the desired position.

Replace the label for better protection against humidity.