Hanna Instruments HI 9813-6 Instruction Manual Download Page 13

13

CLEANING PROCEDURE

For better accuracy in measurements and to ensure a good performance
of the probe, a frequent cleaning is recommended.
For this purpose, soak it in Hanna 

HI 700661 

Cleaning Solution for

5 minutes.

Note

: For particular dirty (as for example protein, oil or grease) see

the “Accessories” section for Hanna specific solutions.

HI 9813-6 only:

CAL-CHECK

TM

 FEATURE / CLEAN & CAL PROCEDURE

The unique Cal-Check

TM

 feature allows the user to check the meter

calibration status at any time. Simply follow the below procedure for
a quick and easy operation.
• Rinse the probe with water.
• Immerse the probe in the 

HI 50021 

check solution.

Note

: If the reading is around pH 4.0, the probe is broken and has

to be replaced.

• Press the 

Check

 key.

• If the meter is calibrated “

Probe is

OK

” message is displayed.

• If cleaning is needed “

Clean Probe

and Calibrate

” message is displayed:

1. immerse the probe in the

HI 700661 

cleaning solution (or

in one of the available specific
solutions - see “Accessories” section)
for 5 minutes, then rinse it with tap water.

2. immerse again in the 

HI 50021 

check solution and press the

CHECK

 key.

3. if the message “

Probe is OK

” is displayed, the meter does not

need calibration.

4. if the message “

Clean Probe and Calibrate

” is displayed

again, calibration is needed (see “pH Calibration” section).

Notes:

 • If it is not possible to calibrate, the probe has to be

replaced with a new one.

• For field applications, it is always recommended to keep a

spare probe handy. When anomalies are not resolved with
simple maintenance, change the probe and recalibrate
the meter.

Summary of Contents for HI 9813-6

Page 1: ...1 Instruction Manual HI 9813 5 HI 9813 6 Portable pH EC TDS C Meters www hannainst com...

Page 2: ...BLE OF CONTENTS TABLE OF CONTENTS HI 9813 5 and HI 9813 6 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to ins...

Page 3: ...ake sure that no damage has occurred during shipment If noticeable damage is found notify your dealer Each meter is supplied with Combination probe with 8 pin DIN connector 1 m 3 3 cable HI 1285 5 pH...

Page 4: ...on key 5 ppm TDS selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 On Off key FUNCTIONAL DESCRIPTION OF HI 9813 5 FUNCTIONAL DESCRIPTION OF HI 9813 5...

Page 5: ...pm TDS range selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 Check key 10 On Off key FUNCTIONAL DESCRIPTION OF HI 9813 6 FUNCTIONAL DESCRIPTION OF...

Page 6: ...a r b i l a C S D T C E e p o l s h g u o r h t t n i o p 1 l a u n a M r e m m i r t e r u t a r e p m e T S D T C E n o i t a s n e p m o C F 2 2 1 o t 2 3 C 0 5 o t 0 m o r f c i t a m o t u A h t...

Page 7: ...simply submerge the tip 4cm 1 of the probe into the sample to be tested Select the pH mode Stir briefly and wait a couple of minutes for the reading to adjust and stabilize The display shows the pH v...

Page 8: ...S mode TAKING TEMPERATURE MEASUREMENTS Submerge the tip 4cm 1 of the probe into the sample to be tested Select the C mode Stir briefly and wait a couple of minutes for the reading to adjust and stabil...

Page 9: ...HI 7009 instead of pH 10 01 PROCEDURE Connect the probe and switch the meter on then press the pH key to display pH measurement Remove the protective cap from the probe rinse and immerse it in the bu...

Page 10: ...4 0 9 5 8 9 0 5 2 2 1 6 0 4 3 8 6 8 9 6 1 0 9 2 8 9 5 5 1 3 1 8 0 4 4 8 6 8 9 6 9 9 8 9 7 9 0 6 0 4 1 9 0 4 4 8 6 8 9 6 7 9 8 7 7 9 5 6 9 4 1 1 1 4 4 8 6 9 9 6 5 9 8 6 7 9 0 7 8 5 1 2 1 4 5 8 6 9 9 6...

Page 11: ...nductivity calibration solution e g HI 7031 into a beaker If possible use plastic beaker to minimize any EMC interference Immerse the probe in the solution Wait for a couple of minutes for thermal equ...

Page 12: ...dous area using a 9V alkaline battery Unscrew the three screws on the rear of the meter remove the battery compartment cover and replace the 9V battery with a new one while observing its polarity Make...

Page 13: ...be is broken and has to be replaced Press the Check key If the meter is calibrated Probe is OK message is displayed If cleaning is needed Clean Probe and Calibrate message is displayed 1 immerse the p...

Page 14: ...ion 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle CONDUCTIVITY TDS CALIBRATION SOLUTIONS HI 7031L 1413 S cm solution 500 mL bottle HI 7442L 1500 ppm mg L solution 500 mL bottle OTHER S...

Page 15: ...follow all necessary steps to correct interferences The glass bulb at the end of the electrode is sensitive to electrostatic discharges Avoid touching this glass bulb at all times During operation ES...

Page 16: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mail tech hannainst com www hannainst com...

Reviews: