Measurement

Use SET/HOLD to select pH mode. Place 

the electrode in the solution to be tested 

while stirring gently. The measurement 

should be taken after the stability tag   

has disappeared. The pH value is displayed 

on the first LCD line and the temperature on the second LCD line.

Note: 

Before taking any pH measurement make sure the tester 

has been calibrated (“CAL“ tag is displayed).

EC/TDS Calibration & Measurement

Calibration Procedure

• From measurement mode, press and hold  /MODE until “CAL” is 

displayed on the second LCD line.

• Place the probe in calibration solution: 

HI70031

 (1413 µS/cm) 

for 

HI98129

 and 

HI70030

 (12.88 mS/cm) for 

HI98130

.

• The LCD displays “OK” for 1 second and the tester returns to 

measurement mode.

Note: 

The “CAL” tag on the LCD indicates that the tester is calibrated.

Measurement

Use SET/HOLD to select either EC or TDS mode. 

Place the electrode in the solution to be tested 

while stirring gently. Use plastic beakers to 

minimize any electromagnetic interferences. 

The measurement should be taken after the 

stability tag   has disappeared. The EC/TDS value is displayed on the 

first LCD line and the temperature on the second LCD line.

Note: 

Before taking any measurement make sure the tester has 

been calibrated (“CAL“ tag is displayed).

Changing EC/TDS Conversion Factor (CONV) & Temperature 

Compensation Coefficient ß (BETA)

• From measurement mode, press and hold  /MODE until “TEMP” 

and the current temperature unit are displayed on the second LCD 

line (e.g. “TEMP °C”).

• Press  /MODE again to show the current conversion factor (e.g. 

“0.50 CONV”) and SET/HOLD to change the conversion factor.

• Press  /MODE to show the current temperature compensation 

coefficient ß (e.g. “% 2.1 BETA”) and SET/HOLD to change the 

temperature compensation coefficient ß.

• Press  /MODE to return to measurement mode.

Maintenance

Please read the information below to ensure the highest possible 

accuracy.

• Fresh buffer solution should be used for each calibration.

• For improved accuracy a two-point calibration is recommended.

• If measurements are taken successively, rinse the probe thoroughly  

in distilled or deionized water to eliminate cross-contamination.

• Clean the electrode monthly using pH and EC cleaning solution. 

Keep the electrode in 

HI7061

 - General purpose cleaning solution 

for 30 minutes and  rinse the probe  thoroughly with water.

• When not in use, add a few drops of 

HI70300

 Storage solution to the 

protective cap. Never store the probe in distilled or deionized water.

Electrode Replacement

Use supplied removal tool (

HI73128

) to replace the pH electrode.

Insert the tool into the probe cavity. 

Rotate the tool counterclockwise.

Pull the electrode out by using the 

other side of the tool. Insert a new 

pH electrode following the above 
instructions in reverse order.

Battery Replacement

Battery life percentage indicator is displayed at power on. If the 

battery level drops below 5%, the   symbol lights up indicating 

that the batteries should be replaced soon.

If the battery level is not adequate to guarantee an accurate reading, 

the tester automatically turns off. 

To replace the batteries, follow the next steps:
1.  Turn OFF the tester.
2.  Remove the four screws on the top of the tester to open the battery 

compartment (figure 1).

3.  Remove the old batteries.
4.  Insert four new 1.5V AA batteries in the battery compartment 

while paying attention to the correct polarity (figure 2).

5.  Close the battery compartment using the four screws.

fig. 1

fig. 2

Note: 

Only use the battery type specified in the manual. Old 

batteries should be disposed in accordance with local regulations.

Accessories

Electrode

HI73127

Replaceable pH electrode

HI73128

Electrode removal tool

Solutions

HI70004P 

pH 4.01 solution, 20 mL sachet (25 pcs.)

HI70006P

pH 6.86 solution, 20 mL sachet (25 pcs.)

HI70007P

pH 7.01 solution, 20 mL sachet (25 pcs.)

HI70009P

pH 9.18 solution, 20 mL sachet (25 pcs.)

HI70010P

pH 10.01 solution, 20 mL sachet (25 pcs.)

HI770710P 

pH 10.01 & 7.01 solution, 20 mL sachet (10 pcs., 5 each)

HI77400P

pH 4.01 & 7.01 solutions, 20 mL sachet (10 pcs., 5 each)

HI70030P

12.88 mS/cm solution, 20 mL sachet (25 pcs.)

HI70031P

1413 µS/cm solution, 20 mL sachet (25 pcs.)

HI70032P

1382 ppm solution, 20 mL sachet (25 pcs.)

HI70038P

6.44 ppt solution, 20 mL sachet (25 pcs.)

HI70442P

1500 ppm solution, 20 mL sachet (25 pcs.)

Electrode Cleaning Solution

HI7061M

General purpose cleaning solution, 230 mL

Electrode Storage Solution

HI70300M

Electrode storage solution, 230 mL

Other Accessories

HI740026P

Replacement 1.5 V batteries (12 pcs.)

Certification

All Hanna Instruments conform to the 

CE European 

Directives.

Disposal of Electrical & Electronic Equipment.

 The 

product should not be treated as household waste. 

Instead hand it over to the appropriate collection point 

for the recycling of electrical and electronic equipment 

which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do not 

dispose of them with other household waste. Hand them over to the 

appropriate collection point for recycling.

Ensuring proper product and battery disposal prevents potential 

negative consequences for the environment and human health. For 

more information, contact your city, your local household waste 
disposal service, the place of purchase or go to www.hannainst.com.

Recommendations for Users

Before using this tester, make sure that it is entirely suitable for your 

specific application and the environment in which it is used. The glass 

bulb at the end of the electrode is sensitive to electrostatic discharges. 

Avoid touching the glass bulb at all times. Any variation introduced 

by the user to the supplied equipment may degrade the tester’s 

performance. For yours and the tester’s safety do not use or store the 

tester in hazardous environments.

Warranty

HI98129

 and 

HI98130

 are warranted for a period of one year against defects 

in workmanship and materials when used for their intended purpose and 

maintained according to instructions. The electrode is warranted for a period 

of six months. This warranty is limited to repair or replacement free of charge. 

Damage due to accidents, misuse, tampering or lack of prescribed maintenance 

is not covered. If service is required, contact your local Hanna Instruments 

Office. If under warranty, report the model number, date of purchase, serial 

number and the nature of the problem. If the repair is not covered by the 

warranty, you will be notified of the charges incurred. 

If the instrument is to be returned to Hanna Instruments Office, first obtain 

a Returned Goods Authorization (RGA) number from the Technical Service 

department and then send it with shipping costs prepaid. When shipping any 

instrument, make sure it is properly packaged for complete protection.

Hanna Instruments reserves the right to modify the design, construction, 

or appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the 

written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, 

Rhode Island, 02895, USA.

IST98129 09/19

HI98129 • HI98130

Waterproof pH, EC/TDS 

& Temperature Testers

INSTRUCTION MANUAL

Summary of Contents for HI 98129

Page 1: ...HI98129 HI98130 Waterproof pH EC TDS Temperature Testers INSTRUCTION MANUAL ...

Page 2: ...ies 4 pcs Instrument quality certificate Instruction manual Each HI98130 is delivered in a cardboard box and is supplied with HI70004 pH 4 01 buffer solution 20 mL sachet 1 pc HI70007 pH 7 01 buffer solution 20 mL sachet 1 pc HI70300NStoragesolutionforpH ORPelectrodes 20mLsachet 1pc HI70030 12880 µS cm conductivity standard 20 mL sachet 1 pc HI70038 6 44 ppt TDS standard 20 mL sachet 1 pc HI700601...

Page 3: ... Range 0 0 to 60 0 C 32 0 to 140 0 F 0 00 to 14 00 pH 0 to 3999 µS cm HI98129 0 00 to 20 00 mS cm HI98130 0 to 2000 ppm HI98129 0 00 to 10 00 ppt HI98130 Resolution 0 1 C 0 1 F 0 01 pH 1 µS cm 1 ppm HI98129 0 01 mS cm 0 01 ppt HI98130 Accuracy 25 C 77 F 0 5 C 1 0 F 0 05 pH 2 f s EC TDS Temperature compensation Automatic EC TDS ß 0 0 to 2 4 ºC Environment 0 to 50 C 32 to 122 F RH 100 TDS factor 0 4...

Page 4: ...nt unit HI98130 7 First LCD line 8 Second LCD line Operational Guide Turning the Tester ON OFF Press and hold MODE All LCD segments will be displayed for a few seconds followed by battery life percentage indicator To turn the tester off from measurement mode press MODE OFF will be displayed on the second LCD line and then the tester will turn off HOLD Mode From measurement mode press and hold SET ...

Page 5: ...le is not valid Forone pointcalibrationusingpH4 01 10 01or9 18buffersolution once the reading has stabilized the tester automatically accepts the calibration point Accepted buffer point and OK 1 message are displayed and then the tester returns to measurement mode For one point calibration using pH 7 01 or 6 86 buffer solution press MODE after calibration point has been accepted The tester will di...

Page 6: ...re on the second LCD line Note Before taking any measurement make sure the tester has been calibrated CAL tag is displayed Changing EC TDS Conversion Factor CONV Temperature Compensation Coefficient ß BETA From measurement mode press and hold MODE until TEMP and the current temperature unit are displayed on the second LCD line e g TEMP C Press MODE again to show the current conversion factor e g 0...

Page 7: ...ert four new 1 5V AA batteries in the battery compartment while paying attention to the correct polarity figure 2 5 Close the battery compartment using the four screws fig 1 fig 2 Note Only use the battery type specified in the manual Old batteries should be disposed in accordance with local regulations Accessories Electrode HI73127 Replaceable pH electrode HI73128 Electrode removal tool Solutions...

Page 8: ...y variation introduced by the user to the supplied equipment may degrade the tester s performance For yours and the tester s safety do not use or store the tester in hazardous environments Warranty HI98129 and HI98130 are warranted for a period of one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions The electrode is war...

Reviews: