Dear Customer, 

Thank you for choosing a Hanna Instruments product. 

Please read this instruction manual carefully before using 

the tester. For more information about Hanna Instruments 

and our products, visit www.hannainst.com or e-mail us at 

[email protected].

For technical support, contact your local Hanna Instruments Office 

or e-mail us at [email protected].

Preliminary Examination

Remove the tester and accessories from the packing material and 

examine it carefully. If you require any further information, please contact 

Hanna Instruments technical support team at [email protected].
Each 

HI98129

 is delivered in a cardboard box and is supplied with:

• 

HI70004

 pH 4.01 buffer solution, 20 mL sachet (1 pc.)

• 

HI70007

 pH 7.01 buffer solution, 20 mL sachet (1 pc.)

• 

HI70300N

 Storage solution for pH & ORP electrodes, 20 mL sachet (1 pc.)

• 

HI70031

 1413 µS/cm conductivity standard, 20 mL sachet (1 pc.)

• 

HI70032

 1382 ppm TDS standard, 20 mL sachet (1 pc.)

• 

HI700601

 pH & ORP electrode cleaning solution, 20 mL sachet (1 pc.)

• 

HI73127

 pH electrode

• 

HI73128

 Electrode removal tool

• 1.5V batteries (4 pcs.)

• Instrument quality certificate

• Instruction manual
Each 

HI98130

 is delivered in a cardboard box and is supplied with:

• 

HI70004

 pH 4.01 buffer solution, 20 mL sachet (1 pc.)

• 

HI70007

 pH 7.01 buffer solution, 20 mL sachet (1 pc.)

• 

HI70300N

 Storage solution for pH & ORP electrodes, 20 mL sachet (1 pc.)

• 

HI70030

 12880 µS/cm conductivity standard, 20 mL sachet (1 pc.)

• 

HI70038

 6.44 ppt TDS standard, 20 mL sachet (1 pc.)

• 

HI700601

 pH & ORP electrode cleaning solution, 20 mL sachet (1 pc.)

• 

HI73127

 pH electrode

• 

HI73128

 Electrode removal tool

• 1.5V batteries (4 pcs.)

• Instrument quality certificate

• Instruction manual

Note: 

Save all packing material until you are sure that the tester 

works correctly. Any damaged or defective item must be returned 

in its original packing material with the supplied accessories.

General Description & Intended Use

HI98129

 & 

HI98130

 are compact pH, EC/TDS and temperature 

testers. They feature a two-buttons operation system and are easy 

to use. The compact and waterproof casing is designed to float if 

accidentally dropped in water. The testers measure temperature in 

both °C and °F. All pH and EC/TDS measurements are temperature 

compensated automatically (ATC). EC/TDS measurements have 

user-selectable EC/TDS conversion factor (CONV) and temperature 

compensation coefficient (ß). The testers display a stability tag ( ) 

that will disappear once the reading has stabilized.

Battery level and low-battery indicator are clearly displayed on 

the LCD to alert the user in the event that low battery power could 

adversely affect readings. 

BEPS (Battery Error Prevention System) Testers will automatically 

shut off if there is not enough power to get an accurate measurement.

pH Probe:

 

HI98129

 & 

HI98130

 are supplied together with 

HI73127

 

pH replaceable electrode with a stainless steel round connector and 

extendable cloth junction. This design has no pins to line up or that 

can break.

Temperature Sensor:

 The stainless steel temperature sensor 

facilitates faster and more accurate temperature measurement.

Specifications

Range

0.0 to 60.0 °C / 32.0 to 140.0 °F

0.00 to 14.00 pH

0 to 3999 µS/cm (

HI98129

)

0.00 to 20.00 mS/cm (

HI98130

)

0 to 2000 ppm (

HI98129

)

0.00 to 10.00 ppt (

HI98130

)

Resolution

0.1 °C (0.1 °F)

0.01 pH

1 µS/cm ; 1 ppm (

HI98129

)

0.01 mS/cm ; 0.01 ppt (

HI98130

)

Accuracy

(@25 °C / 77 °F)

±0.5 °C (±1.0 °F)

±0.05 pH

±2% f.s. (EC/TDS)

Temperature 

compensation

Automatic 

EC/TDS: ß=0.0 to 2.4% / ºC 

Environment

0 to 50 °C (32 to 122 °F)

RH 100%

TDS factor

0.45 to 1.00 (CONV)

Calibration

pH: automatic, one or two-point with two 

sets of standard buffers (pH 4.01 / 7.01 / 

10.01 or pH 4.01 / 6.86 / 9.18)

EC/TDS: automatic, one-point calibration

EC/TDS Calibration 

Solutions

HI98129

HI98130

HI70031

 (1413 µS/cm)

HI70032

 (1382 ppm; CONV=0.50)

HI70442

 (1500 ppm; CONV=0.70)

HI70030

 (12.88 mS/cm)

HI70038 

(6.44 ppt; CONV=0.50

or 9.02 ppt; CONV=0.70)

Electrode

HI73127

 pH electrode (included) 

Battery type

1.5V (4 pcs.)

Battery life

Approx. 100 hours

Auto-off

After 8 minutes of non-use

Dimensions

171 x 41 x 26 mm (6.7 x 1.6 x 1.0”)

Weight

93 g (3.4 oz.)

Functional Description

Liquid Crystal 

Display
SET/HOLD 

button

Battery 

compartment

/MODE 

button

EC/TDS 

electrode

Temperature 

sensor 

(inside)

HI73127 

pH electrode

LCD Display

1.  Automatic Temperature Compensation (ATC) indicator

2.  Stability tag

3.  Battery life percentage indicator

4.  Low battery indicator

5.  Measurement unit (

HI98129

)

6.  Measurement unit (

HI98130

)

7.  First LCD line

8.  Second LCD line

Operational Guide

Turning the Tester ON / OFF

Press and hold  /MODE. All LCD segments will be displayed for a 

few seconds followed by battery life percentage indicator. To turn the 

tester off, from measurement mode, press  /MODE. “OFF” will be 

displayed on the second LCD line and then the tester will turn off.

HOLD Mode 

From measurement mode, press and hold SET/HOLD until “HOLD” is 

displayed. pH, EC/TDS or temperature readings will be frozen on the 

LCD. Press either buttons to resume measurement mode.

Setup

Setup mode allows the selection of temperature unit and pH buffer set. To 

enter the setup mode, select pH mode and press  /MODE until “TEMP” 

and the current temperature unit are displayed (e.g. “TEMP °C”). Then:

• Press SET/HOLD to select temperature unit. After the temperature 

unit has been selected, press  /MODE to select the buffer set. To 

return to measurement mode, press  /MODE twice.

• After setting the temperature unit, the tester will display the current 

buffer set: “pH7.01 BUFF” (for pH 4.01/7.01/10.01) or “pH6.86 

BUFF” (for NIST pH 4.01/6.86/9.18). Press SET/HOLD to change 

the buffer set and press  /MODE to resume measurement mode.

pH Calibration & Measurement

It is recommended to calibrate the tester frequently, especially if high 

accuracy is required. More frequent calibrations may be required depending 

on the type of sample being tested. The tester should be recalibrated:

• whenever the pH electrode is replaced

• at least once a month

• after testing aggressive samples

From measurement mode, press and hold  /MODE until “CAL” is 

displayed. The tester enters calibration mode and displays “pH 7.01 

USE” or “pH 6.86 USE”, if NIST buffer was selected. Buffer value 

will be displayed on the first LCD line and “REC” message will be 

displayed on the second line. If used buffer was not valid, “USE” 

message will be displayed for 12 seconds and replaced by “WRNG” 

indicating measured sample is not valid.

• For one-point calibration using pH 4.01, 10.01 or 9.18 buffer solution, 

once the reading has stabilized, the tester automatically accepts the 

calibration point. Accepted buffer point and “OK 1” message are 

displayed and then the tester returns to measurement mode.

• For one-point calibration using pH 7.01 (or 6.86) buffer solution, 

press  /MODE after calibration point has been accepted. The 

tester will display “pH 7.01” (or “pH 6.86”) and “OK 1” and then  

return to measurement mode.

• For a two-point calibration using pH 7.01 (or pH 6.86) buffer solution, 

after the first calibration point has been accepted, “pH 4.01 USE” will 

be displayed for 12 seconds (unless a valid buffer is recognized). 

• If valid buffer solution is recognized (pH 4.01, 10.01 or 9.18), 

the calibration point is accepted and the accepted value and “OK 

2” are displayed. The tester then returns to measurement mode.

• If no valid buffer solution is recognized, “WRNG” message is displayed.

Note: 

When the calibration procedure is completed, the “CAL” 

tag is turned on.

Exiting Calibration & Resetting Default Values

• When in calibration mode, it is possible to exit calibration 

procedure by pressing  /MODE, before first-point calibration 

has been accepted. The tester displays “ESC” and returns to 

measurement mode and last calibrated data.

• When in calibration mode, it is possible  to clear a previous 

calibration and return to default values by pressing SET/HOLD, 

before the first calibration point has been accepted. The tester 

displays  “CLR” on the second LCD line, “CAL” tag dissappears and 

tester returns to default calibration.

Summary of Contents for HI 98129

Page 1: ...HI98129 HI98130 Waterproof pH EC TDS Temperature Testers INSTRUCTION MANUAL ...

Page 2: ...ies 4 pcs Instrument quality certificate Instruction manual Each HI98130 is delivered in a cardboard box and is supplied with HI70004 pH 4 01 buffer solution 20 mL sachet 1 pc HI70007 pH 7 01 buffer solution 20 mL sachet 1 pc HI70300NStoragesolutionforpH ORPelectrodes 20mLsachet 1pc HI70030 12880 µS cm conductivity standard 20 mL sachet 1 pc HI70038 6 44 ppt TDS standard 20 mL sachet 1 pc HI700601...

Page 3: ... Range 0 0 to 60 0 C 32 0 to 140 0 F 0 00 to 14 00 pH 0 to 3999 µS cm HI98129 0 00 to 20 00 mS cm HI98130 0 to 2000 ppm HI98129 0 00 to 10 00 ppt HI98130 Resolution 0 1 C 0 1 F 0 01 pH 1 µS cm 1 ppm HI98129 0 01 mS cm 0 01 ppt HI98130 Accuracy 25 C 77 F 0 5 C 1 0 F 0 05 pH 2 f s EC TDS Temperature compensation Automatic EC TDS ß 0 0 to 2 4 ºC Environment 0 to 50 C 32 to 122 F RH 100 TDS factor 0 4...

Page 4: ...nt unit HI98130 7 First LCD line 8 Second LCD line Operational Guide Turning the Tester ON OFF Press and hold MODE All LCD segments will be displayed for a few seconds followed by battery life percentage indicator To turn the tester off from measurement mode press MODE OFF will be displayed on the second LCD line and then the tester will turn off HOLD Mode From measurement mode press and hold SET ...

Page 5: ...le is not valid Forone pointcalibrationusingpH4 01 10 01or9 18buffersolution once the reading has stabilized the tester automatically accepts the calibration point Accepted buffer point and OK 1 message are displayed and then the tester returns to measurement mode For one point calibration using pH 7 01 or 6 86 buffer solution press MODE after calibration point has been accepted The tester will di...

Page 6: ...re on the second LCD line Note Before taking any measurement make sure the tester has been calibrated CAL tag is displayed Changing EC TDS Conversion Factor CONV Temperature Compensation Coefficient ß BETA From measurement mode press and hold MODE until TEMP and the current temperature unit are displayed on the second LCD line e g TEMP C Press MODE again to show the current conversion factor e g 0...

Page 7: ...ert four new 1 5V AA batteries in the battery compartment while paying attention to the correct polarity figure 2 5 Close the battery compartment using the four screws fig 1 fig 2 Note Only use the battery type specified in the manual Old batteries should be disposed in accordance with local regulations Accessories Electrode HI73127 Replaceable pH electrode HI73128 Electrode removal tool Solutions...

Page 8: ...y variation introduced by the user to the supplied equipment may degrade the tester s performance For yours and the tester s safety do not use or store the tester in hazardous environments Warranty HI98129 and HI98130 are warranted for a period of one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions The electrode is war...