Hanna Instruments HI 9810 Instruction Manual Download Page 4

7

6

TAKING EC/TDS MEASUREMENTS

• Immerse the tip of the electrode (4cm/1½") into the

sample to be tested. If possible, use plastic beakers
or containers to minimize any EMC interference.

• Tap the electrode lightly on the bottom of the

beaker to remove any air bubbles which may be
trapped inside the tip.

• Select the appropriate measurement range (EC or TDS).

• Wait for 1-2 minutes for the temperature sensor to attain thermal

equilibrium. The display will then show the measurement auto-
matically temperature compensated for temperature with the
appropriate indication among the following:

µS 

symbol indicates the meter is in

µS/cm, EC mode

No symbol indicates the meter is in TDS

mode.

Note:

 If the case of an erroneous reading, the "E" indication will

appear on the right of the display.

AFTER MEASUREMENTS

After measurements have been completed, the instrument should be
switched off and the probe cleaned and covered with the protective
cap.

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

INITIAL PREPARATION

Each meter is supplied complete with a 9V battery. Slide off the
battery compartment cover on the back of the meter (see page 16),
install the battery while paying attention to its polarity.
Connect the probe to the DIN socket on the
top of the meter by aligning the pins with
the socket and pushing in the plug.
Always remove the electrode protective cap before taking any mea-
surements.
Make sure the meter has been calibrated before taking any measure-
ments (see pages 8 and 9 for calibration
procedures).
Turn the meter on by pressing the ON/OFF key.

TAKING  PH MEASUREMENTS

If the electrode has been left dry, soak the tip in a pH 7 or pH 4
buffer solution for a few minutes to reactivate it.

• To take a pH measurement simply submerge the tip

(4cm/1½") of the combination pH/EC/TDS electrode
into the sample to be tested.

• Select the pH mode .

• Shake the electrode briefly while sub-

merged and allow a couple of minutes
for the electrode to adjust and stabilize.
The display will show the pH value.

• If measurements are taken in different samples successively, it is

recommended to rinse (clean) the electrode thoroughly to elimi-
nate cross-contamination. After cleaning, it is recommended to
rinse the electrode with some of the sample to be measured.

Summary of Contents for HI 9810

Page 1: ...Instruction Manual HI 9810 Portable pH EC TDS Meter h t t p w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Page 2: ...rature and for this reason measurements are carried out with reference to a standard temperature of 25 C If the solution measured has a different tem perature than 25 C compensation must be performed...

Page 3: ...with a of 2 C Electrode HI 1285 combination pH EC TDS electrode with temperature sensor and 1 m 3 3 cable included Battery Type 9 Volt 150 hours of continuous use Environment 0to50 C 32to122 F max 95...

Page 4: ...N Each meter is supplied complete with a 9V battery Slide off the battery compartment cover on the back of the meter see page 16 install the battery while paying attention to its polarity Connect the...

Page 5: ...chemicals d Where extreme accuracy is required PREPARATION Pour small quantities of pH 7 01 HI 7007 or HI 8007 orpH4 01 HI 7004orHI 8004 or pH 10 01 HI 7010 or HI 8010 solution into a clean beaker To...

Page 6: ...gs are usually normalized at 25 C HI 9810 automatically compensates for temperature differences with a built in temperature sensor circuitry With this compensation the display shows the readings at 25...

Page 7: ...e electrode and recalibrate the meter to see if the problem is alleviated ELECTRODE CONDITIONING ELECTRODE CONDITIONING ELECTRODE CONDITIONING ELECTRODE CONDITIONING ELECTRODE CONDITIONING AND MAINTEN...

Page 8: ...1 Rugged carrying case ACCESSORIES ACCESSORIES ACCESSORIES ACCESSORIES ACCESSORIES pH CALIBRATION SOLUTIONS HI 70004P pH 4 01 Buffer Sachets 25 x 20 mL HI 7004M pH 4 01 Buffer Solution 230 mL HI 7004L...

Page 9: ...EMICAL TEST KITS CHLORINE METERS CONDUCTIVITY TDS METERS DISSOLVED OXYGEN METERS HYGROMETERS ION SPECIFIC METERS Colorimeters MAGNETIC STIRRERS Na NaCl METERS pH ORP Na ELECTRODES PROBES DO S cm RH T...

Page 10: ...warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions The electrodes and the probes are warranted for a per...

Page 11: ...plications The reference literature currently covers areas such as Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others New reference material is constantly b...

Reviews: