background image

To save battery, the instrument shuts down after 10 minutes of non-use
in 

measurement mode and after 1 hour of non-use in calibration mode.

If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking “

ZERO

means that a new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup
and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken
anymore, the instrument shows “

dEAd bAtt

” and turns off.

To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing 

ON/OFF

.

• Turn the instrument upside down and remove the battery cover by

turning it counterclockwise.

• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.

INTERFERENCES

• Alkalinity: above 1,000 mg/L CaCO

3

 if present

as bicarbonate (HCO

3

 sample pH 

<

 8.3);

above 25 mg/L CaCO

3

 if present as carbonate

(CO

3

2-

, sample pH > 9.0). In both cases, it

will not reliably develop the full amount of
color or it may rapidly fade (negative error).
To resolve this, neutralize the sample with
diluted HCl.

• Acidity: above 150 mg/L CaCO

3

. May not

reliably develop the full amout of color or it
may rapidly fade (negative error). To resolve
this, neutralize the sample with diluted NaOH.

• Hardness: in case of water with hardness

greater than 500 mg/L CaCO

3

, shake the

sample for approximately 2 minutes after
adding the powder reagent.

• Bromine (Br

2

),  Chloride dioxide (ClO

2

), 

Iodine

(I

2

), Oxidized Manganese and Chromium,

Ozone (O

3

): positive error.

R

EAGENT SETS

HI 95762-01

Reagents for 100 tests

HI 95762-03

Reagents for 300 tests

OTHER ACCESSORIES

HI 96762-11

CAL CHECK™ Standard Cuvettes (1 set)

HI 721310

9V battery (10 pcs)

HI 731318

Cloth for wiping cuvettes (4 pcs)

HI 731331

Glass cuvettes (4 pcs)

HI 731335

Caps for cuvettes (4 pcs)

HI 93703-50

Cuvette cleaning solution (230 mL).

FACTORY CALIBRATION RESTORE

It is possible to delete the calibration and
restore factory calibration.

1•

Press 

GLP/

 to enter 

GLP mode.

2•

Press 

READ/ /TIMER

 to enter in the

factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.

3•

Press 

ZERO/CFM

 to restore the factory

calibration or press 

GLP/

 again to

abort factory calibration restore.

4•

The instrument briefly indicates 

“donE”

upon restoration of factory calibration prior
to returning to 

measurement mode.

LAST CALIBRATION DATE

1•

Press 

GLP/

 to enter 

GLP mode. The

calibration month and day will appear on
the main display and the year on the
secondary display.

2•

If no calibration was performed, the factory
calibration message, 

“F.CAL”

 will appear

on the main display and the instrument
returns to 

measurement mode after three

seconds.

11•

Then the date of last calibration (e.g.:

01.08.2009

”) appears on the display, or

01.01.2009

” if the factory calibration

was selected before. In both cases the year
number is blinking, ready for date input.

12•

Press 

GLP/

 to edit the desired year

(2009-2099). If the key is kept pressed,
the year number is automatically increased.

13•

When the correct year has been set, press

ZERO/CFM

 or 

READ/ /TIMER

 to confirm.

Now the display will show the month
blinking.

14•

Press 

GLP/

 to edit the desired month

(01-12). If the key is kept pressed, the
month number is automatically increased.

15•

When the correct month has been set,
press 

ZERO/CFM

 or 

READ/ /TIMER

 to

confirm. Now the display will show the
day blinking.

16•

Press 

GLP/

 to edit the desired day

(01-31). If the key is kept pressed, the
day number is automatically increased.

Note:

It is possible to change the editing from
day to year and to month by pressing

READ/ /TIMER.

17•

Press 

ZERO/CFM

 

to save the calibration

date.

18•

The instrument displays “

Stor

” for one

second and the calibration is saved.

19•

The instrument will return automatically
to 

measurement mode by displaying dashes

on the LCD.

Calibration 

9•

At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications
as reported on the 

CAL CHECK™

Standard Certificate. If the value is found
out of specifications, please check that the
cuvettes are free of fingerprints, oil or dirt
and repeat validation. If results are still
found out of specifications then recalibrate
the instrument.

GLP

Battery management

Warranty

HI 96762

 is warranted for two years against defects in workmanship and

materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause  unacceptable interferences to other electronic equipments, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in  hazardous environments.

Hanna Instruments 

reserves the right to modify the

design, construction and appearance of its products

without advance notice.

CALIBRATION

Note:

It is possible to interrupt the calibration
procedure at any time by pressing 

CAL

CHECK

 or 

ON/OFF

 keys.

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.

3•

Press and hold 

CAL CHECK

 for three

seconds to enter 

calibration mode. The

display will show “

CAL

” during calibration

procedure. The blinking “

ZERO

” asks for

instrument zeroing.

4•

Place the 

CAL CHECK™

 Standard

HI 96762-11

 Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

5•

Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

6•

After a few seconds the display will show

-0.0-

”. The meter is now zeroed and

ready for calibration. The blinking “

READ

asks for reading calibration standard.

7•

Remove the cuvette.

8•

Place the 

CAL CHECK™

 Standard

HI 96762-11

 Cuvette B into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

9•

Press 

READ/ /TIMER

 and the lamp,

cuvette and detector icons will appear on
the display, depending on the measurement
phase.

10•

The instrument will show for three seconds
the 

CAL CHECK™

 standard value.

Note:

If the display shows “

STD HIGH

”, the

standard value was too high. If the
display shows “

STD LOW

”, the standard

value was too low. Verify that both

CAL CHECK™

 Standard 

HI 96762-11

Cuvettes, A and B are free from fingerprints
or dirt and that they are inserted correctly.

IST96762   01/11

9

3

4

5-6

8

9-10

11-13

 or

 or

16

17

18

Last

Calibration

Date

 

In the 

GLP mode, the last calibration date can be verified and the factory

calibration can be restored.

1

2

Factory

Calibration

Restore

 

2

3-4

Accessories:

3

Warning:

 do not validate or calibrate the instrument with standard

solutions other than the Hanna 

CAL CHECK™

 Standards, otherwise

erroneous results will be obtained.

Use the Hanna CAL CHECK™ cuvettes (see
“Accessories”) to validate or calibrate instruments.

Validation and Calibration procedures

VALIDATION

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.

3•

Place the 

CAL CHECK™

 Standard

HI 96762-11

 Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

4•

Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

5•

After a few seconds the display will show

-0.0-

”. The meter is now zeroed and

ready for validation.

6•

Remove the cuvette.

7•

Place the 

CAL CHECK™

 Standard

HI 96762-11

 Cuvette B into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

8•

Press 

CAL CHECK

 key and the lamp,

cuvette and detector icons together with

CAL CHECK

” will appear on the display,

depending on the measurement phase.

Validation 

4-5

7

8

 or

11

14-15

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com 

Reviews: