background image

11•

Press and hold 

READ/ /TIMER

 for three

seconds

.

 The display will show the countdown

prior to measurement (not for Hardness).

The beeper is playing a beep at the end of

countdown period.

Alternatively, wait for:

Free Chlorine:

1 minute

Total Chlorine:

2 minutes and 30 seconds

Iron LR:

30 seconds

Then press 

READ/ /TIMER

.

For pH and Hardness press 

READ/ /

TIMER

 directly.

In all cases the lamp, cuvette and detector

icons will appear on the display, depending

on the measurement phase.

12•

The instrument directly displays the pH

measured value or the concentration in

mg/L of free chlorine, total chlorine or iron

on the LCD, depending on the selected

parameter.

For Hardness the instrument will display the concentration of

Calcium in mg/L CaCO

3

 (together with ‘’C’’). Press 

READ/ /

TIMER

 again, the lamp, cuvette and detector icons will appear on

the display, depending on the measurement phase. The instrument

will display the total hardness concentration in mg/L  CaCO

3

(together with ‘’t’’).

INTERFERENCES:

• Chlorine: Bromine, Iodine, Chlorine dioxide. Ozone and Oxidized forms

of Manganese and Chromium. Alkalinity above 250 mg/L CaCO

or

acidity above 150 mg/L CaCO

3

 will not reliably develop the full

amount of color or it may rapidly fade. To resolve this, neutralize the

sample with diluted HCl or NaOH. In case of water with hardness

greater than 500 mg/L CaCO

3

, shake the sample for approximately 2

minutes after adding the powder reagent.

• Iron LR: Cadmium above 4.0 mg/L, C above 1.2 mg/L,

Copper above 0.6 mg/L, Manganese above 50.0 mg/L, Molybdenum

4.0 mg/L, Nitrite ion above 0.8 mg/L, Ch above 0.25

mg/L, Cobalt above 0.05 mg/L, Cyanide above 2.8 mg/L, Mercury

above 0.4 mg/L, Nickel above 1.0 mg/L.

• Total Hardness: Excessive amounts of heavy metals.

Note

: If the sample is very acidic, some extra drops of 

HI 93735B

 buffer

reagent may be added.

Use the Hanna CAL CHECK™ cuvettes (see
“Accessories”) to validate or calibrate instruments.

Validation and Calibration procedures

VALIDATION
Note:

 The validation is performed only for the

selected parameter. For full validation of

the instrument, the following procedure

must be performed for each parameter.

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

3•

Place the 

CAL CHECK™

 Standard Cuvette  A

into the cuvette holder and ensure that the

notch on the cap is positioned securely into the

groove.

4•

Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the display,

depending on the measurement phase.

5•

After a few seconds the display will show

-0.0-

”. The meter is now zeroed and

ready for validation.

Validation 

12

8

Calibration 

CALIBRATION

Note:

 It is possible to interrupt the calibration

procedure at any time by pressing

CAL CHECK

 or 

ON/OFF

 keys.

When calibrating, only the selected range

is affected.

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

3•

To change the range, simply press

RANGE/GLP

/ .

4•

Press and hold 

CAL CHECK 

for three seconds

to enter 

calibration mode. The display will

show 

“CAL”

 during calibration procedure. The

blinking 

“ZERO”

 asks for instrument zeroing.

5•

Place the 

CAL CHECK™

 Standard Cuvette A

into the cuvette holder and ensure that the

notch on the cap is positioned securely into the

groove.

6•

Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the display,

depending on the measurement phase.

7•

After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and

ready for calibration. The blinking 

“READ”

asks for reading calibration standard.

8•

Remove the cuvette.

9•

Place the specific 

CAL CHECK

TM

 Standard

Cuvette B into the cuvette holder, for:
pH: 

B, HI 96710-11

Free Chlorine: 

B

HI 96701-11

Total Chlorine: 

B

HI 96711-11

Hardness: 

B, HI 96719-11

Iron LR: 

B

HI 96746-11

Ensure that the notch on the cap is positioned
securely into the groove.

4

5

9

6•

Remove the cuvette.

7•

Place the specific 

CAL CHECK

TM

 Standard

Cuvette B into the cuvette holder, for:
pH: 

B

HI 96710-11

Free Chlorine: 

B

HI 96701-11

Total Chlorine: 

B

HI 96711-11

Hardness: 

B

HI 96719-11

Iron LR: 

B

HI 96746-11

Ensure that the notch on the cap is positioned
securely into the groove.

8•

Press 

CAL CHECK

 key and the lamp,

cuvette and detector icons together with

CAL CHECK

” will appear on the display,

depending on the measurement phase.

9•

At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as
reported on the 

CAL CHECK™

 Standard

Certificate. If the value is found out of
specifications, please check that the cuvettes
are free of fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the instrument.

LAST CALIBRATION DATE

1•

Press and hold 

RANGE/GLP/

 for three

seconds to enter 

GLP mode. The calibration

month and day will appear on the main
display and the year on the secondary display.

2•

If no calibration was performed, the factory
calibration message, 

“F.CAL”

 will appear on

the main display and the instrument returns
to 

measurement mode after three seconds.

10•

Press 

READ/ /TIMER

 and the lamp,

cuvette and detector icons will appear on
the display, depending on the measurement
phase.

11•

The instrument will show for three seconds
the 

CAL CHECK™

 standard value.

Note:

If the display shows 

“STD HIGH”

, the

standard value was too high. If the display
shows 

“STD LOW”

, the standard  value

was too low. Verify that both 

CAL

CHECK™

 Standard Cuvettes, A and B

are free from fingerprints or dirt and that
they are inserted correctly.

12•

Then the date of last calibration (e.g.:

“01.08.2009”

) appears on the display, or

“01.01.2009”

 if the factory calibration

was selected before. In both cases the year
number is blinking, ready for date input.

13•

Press 

RANGE/GLP/

 to edit the desired

year (2009-2099). If the key is kept pressed,
the year number is automatically increased.

14•

When the correct year has been set, press

ZERO/CFM

 or 

READ/ /TIMER

 to confirm.

Now the display will show the month blinking.

15•

Press 

RANGE/GLP/

 to edit the desired

month (01-12). If the key is kept pressed,
the month number is automatically increased.

16•

When the correct month has been set, press

ZERO/CFM

 or 

READ/ /TIMER

 to confirm.

Now the display will show the day blinking.

17•

Press 

RANGE/GLP/

 to edit the desired

day (01-31). If the key is kept pressed,
the day number is automatically increased.

Note:

It is possible to change the editing from
day to year and to month by pressing

READ/ /TIMER.

18•

Press 

ZERO/CFM

 to save the calibration date.

19•

The instrument displays “

Stor

” for one

second and the calibration is saved.

20•

The instrument will return automatically
to 

measurement mode by displaying dashes

on the LCD.

GLP

12-14

 or

15-16

 or

17

18

19

Last

Calibration

Date

 

In 

GLP mode, the last calibration date can be verified and the factory

calibration can be restored.

1

FACTORY CALIBRATION RESTORE

It is possible to delete the calibration and restore
factory calibration.

1•

Press and hold 

RANGE/GLP/

 for three

seconds to enter 

GLP mode.

2•

Press 

READ/ /TIMER

 to enter in the

factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.

3•

Press 

ZERO/CFM

 to restore the factory

calibration or press 

RANGE/GLP/

again to abort factory calibration restore.

4•

The instrument briefly indicates 

“donE”

upon restoration of factory calibration prior
to returning to 

measurement mode.

Battery management

To save the battery, the instrument shuts down after 10 minutes of non-
use in 

measurement mode and after 1 hour of non-use in calibration

mode.
If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking 

“ZERO”

means that a new zero has to be performed.
One fresh battery lasts for arround 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup
and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any
more, the instrument shows 

“dEAd bAtt”

 and turns off.

To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing 

ON/OFF

.

• Turn the instrument upside down and remove the battery cover by

turning it counterclockwise.

• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.

Factory

Calibration

Restore

 

R

EAGENT SETS

HI 93701-01

Reagents for 100 Free Chlorine tests

HI 93701-03

Reagents for 300 Free Chlorine tests

HI 93710-01

Reagents for 100 pH tests

HI 93710-03

Reagents for 300 pH tests

HI 93711-01

Reagents for 100 Total Chlorine tests

HI 93711-03

Reagents for 300 Total Chlorine tests

HI 93719-01

Reagents for 100 Hardness tests

HI 93719-03

Reagents for 300 Hardness tests

HI 93746-01

Reagents for 100 Iron Low Range tests

HI 93746-03

Reagents for 300 Iron Low Range tests

OTHER ACCESSORIES

HI 96701-11 CAL CHECK

TM

 Standard Cuvettes for Free Chlorine (1 set)

HI 96710-11 CAL CHECK

TM

 Standard Cuvettes for pH (1 set)

HI 96711-11 CAL CHECK

TM

 Standard Cuvettes for Total Chlorine (1 set)

HI 96719-11 CAL CHECK

TM

 Standard Cuvettes for Hardness (1 set)

(equivalent with 1.00 mg/L Mg Hardness)

HI 96746-11 CAL CHECK

TM

 Standard Cuvettes for Iron LR (1 set)

HI 721310

9V battery (10 pcs.)

HI 731318

Cloth for wiping cuvettes (4 pcs.)

HI 731331

Glass cuvettes (4 pcs.)

HI 731335

Caps for cuvettes

HI 93703-50

Cuvette cleaning solution (230 mL)

Warranty

HI 96745

 is warranted for two years against defects in workmanship and

materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Depart-
ment and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause  unacceptable interferences to other electronic equipment, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in  hazardous environments.

Hanna Instruments 

reserves the right to modify the

design, construction or appearance of its products

without advance notice.

IST96745   01/13

Accessories

2

10-11

6-7

7

or

9

3

4-5

2

11

3-4

Warning:

 do not validate or calibrate the instrument with standard

solutions other than the Hanna 

CAL CHECK™

 Standards, otherwise

erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at
room temperature (18 to 25°C; 64.5 to 77.0°F).

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com 

Reviews: