Hanna Instruments HI 96741C Instruction Manual Download Page 9

16

17

• Place the cuvette into the cuvette holder and

ensure that the notch on the cap is positioned
securely into the groove.

• Press 

ZERO/CFM

 and the lamp, cuvette and

detector icons will appear on the display,
depending on the measurement phase.

• For Total Hardness: Place the ZERO sample

into the holder and ensure that the notch on
the cap is positioned securely into the groove.
Press 

ZERO/CFM

 and the lamp, cuvette and

detector icons will appear on the display,
depending on the measurement phase. After a
few seconds the display will show 

   “-0.0-”

.

Remove the ZERO sample and insert the READ1
sample into the instrument. Press and hold

READ/ /TIMER

 for three seconds

.

 The display

will show the countdown prior to measurement.
The beeper is playing a beep at the end of
countdown period. Alternatively, wait for 30
seconds. The lamp, cuvette and detector icons
will appear on the display, depending on the
measurement phase, then the instrument will
display the level of Magnesium hardness in
mg/L  CaCO

3

 (togheter with “n”).

• After a few seconds the display will show 

“-

0.0-”

. The meter is now zeroed and ready for

measurement.

• Remove the cuvette.
• Add the specific test reagent for each parameter:

Total hardness: the third cuvette (nothing is
added)
Iron LR: Fill one graduated mixing cylinder up
to the 25 mL mark with the sample.
Add the content of one packet of 

HI 93741-0

reagent, close the cylinder and shake well for
30 seconds.
Fill a cuvette with 10 mL of the reacted
sample up to the mark and replace the cap.

To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.

Important note:

 Free and Total Chlorine have to be measured separately with fresh unreacted

samples if both values are requested.

MEASUREMENT PROCEDURE

• Turn the meter on by pressing 

ON/OFF

. The

display briefly shows all tags on.

• When the beeper sounds briefly and the LCD

displays dashes, the meter is ready. The
blinking 

“ZERO”

 indicates that the instrument

needs to be zeroed first.

• Pay attention to the selected range.

To change the range, simply press 

RANGE/

GLP/

.

• For Total Hardness: Fill a graduated beaker

up to the 50mL mark with the sample. Add
0.5 mL of   

HI 93719A-0

 Calcium And

Magnesium Reagent indicator solution and
mix. Add 0.5 mL of 

HI 93719B-0

 Alkali

solution for Calcium and Magnesium and mix.
Fill three cuvettes with 10mL of sample each.
Add 1 drop of 

HI 93719C-0 EDTA

 solution

to one cuvette, replace the cap and swirl the
solution. This is the ZERO sample. Add 1 drop
of 

HI 93719D-0 EGTA

 solution to the second

cuvette, replace the cap and swirl the solution.
This is the READ1 sample.

For Iron LR: Fill one graduated mixing cylinder
up to the 25 mL mark with deionized water.
Add the content of one packet of 

HI 93741-0

reagent, close the cylinder and shake well for
30 seconds. This is the blank.

Fill a cuvette with 10 mL of the blank up to
the mark and replace the cap.

#1

#2

#3

#1

#2

or

Summary of Contents for HI 96741C

Page 1: ...1 w w w h a n n a i n s t c o m Instruction Manual HI 96741C Total Hardness and Iron LR ISM...

Page 2: ...LP 23 BATTERY MANAGEMENT 24 BATTERY REPLACEMENT 24 ACCESSORIES 25 WARRANTY 26 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the c...

Page 3: ...ive positive locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell It is designed to fit a cuvette with a larger neck making it easier to...

Page 4: ...4 70 mg L Iron LR 0 00 to 1 60 mg L Resolution 0 01 mg L Total Hardness 0 01 mg L Iron LR Accuracy Mg Hardness 0 11 mg L 5 of reading 25 C Ca Hardness 0 11 mg L 5 of reading 25 C Iron LR 0 01 mg L 8...

Page 5: ...oid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvette walls with hands Furthermore in order to maintain the same conditions during...

Page 6: ...t contain any debris Inverted cuvettes The sample and the zero cuvette are inverted Zero A zero reading was not taken Follow the instructions of the measurement procedure for zeroing the meter b on sa...

Page 7: ...tes that the battery is dead and must be replaced Once this indication is displayed normal operation of the instrument will be interrupted Change the battery and restart the meter Under range A blinki...

Page 8: ...s This would corrupt the reading Each time the cuvette is used the cap must be tightened to the same degree Whenever the cuvette is placed into the measurement cell it must be dry outside and complete...

Page 9: ...seconds Fill a cuvette with 10 mL of the reacted sample up to the mark and replace the cap To compensate the meter for the sample turbidity or color the measurement takes place in two phases First th...

Page 10: ...ve 4 0 mg L Nitrite ion above 0 8 mg L Chromium3 above 0 25 mg L Cobalt above 0 05 mg L Cyanide above 2 8 mg L Mercury above 0 4 mg L Nickel above 1 0 mg L Sample pH should be between 3 and 4 to avoid...

Page 11: ...ate input CALIBRATION PROCEDURE Note It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON OFF keys Warning Do not calibrate the instrument with standard solutions...

Page 12: ...RY CALIBRATION RESTORE In GLP mode the last user calibration date can be verified and the factory calibration can be restored LAST CALIBRATION DATE GLP Press RANGE GLP to edit the desired year 2009 20...

Page 13: ...ue is displayed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The rema...

Page 14: ...intained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If se...

Page 15: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mailtech hannainst com www hannainst com...

Reviews: