background image

6

7

PRINCIPLE OF OPERATION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and
matter. When a light beam crosses a substance, some of the radiation may be absorbed by
atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length
through the matter and on the 

physical

-chemical characteristics of the substance according to the

Lambert-Beer Law:

-log

 

I

/

I

o

 = 

ε

λ

 

c d

or

A  

ε

λ

 

c d

Where:

-log

 

I

/

I

o

=

Absorbance (A)

I

o

=

intensity of incident light beam

I

=

intensity of light beam after absorption

ε

λ

=

molar extinction coefficient at wavelength

 

λ

c

=

molar concentration of the substance

d

=

optical path through the substance

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the
other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound
from a specific chemical reaction between sample and reagents. Given that the absorption of a
compound strictly depends on the wavelength of the incident light beam, a narrow spectral
bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's 

HI 96

 series colorimeters is based on special subminiature

tungsten lamps and narrow-band interference filters to guarantee both high performance and
reliable results.

HI 96 series block diagram (optical layout)

SPECIFICATIONS

PRECISION AND ACCURACY

Range

Mg Hardness 0.00 to 2.00 mg/L
Ca Hardness 0.00 to 2.70 mg/L
Total Hardness

 

0.00 to 4.70 mg/L

Iron LR 0.00 to 1.60 mg/L

Resolution

0.01 mg/L Total Hardness
0.01 mg/L Iron LR

Accuracy

Mg Hardness ±0.11 mg/L ±5% of reading @ 25°C
Ca Hardness ±0.11 mg/L ±5% of reading @ 25°C
Iron LR ±0.01 mg/L ±8% of reading @ 25°C

Typical EMC Deviation

±0.02 mg/L Mg Hardness
±0.02 mg/L Ca Hardness
±0.01 mg/L Iron LR

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525nm.

Method

For Total Hardness: Adaptation of the 

Standard Methods for the Examination

of Water and Wastewater, 18

th

 Edition, 

colorimetric method. The reaction

between Mg/Ca and reagents causes a violet tint in the sample.
For Iron LR: Adaptation of the TPTZ method. The reaction between iron and
the reagent causes a violet tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in 

measurement mode

;

after 1 hour of non-use in 

calibration mode

;

with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7")

Weight

360 g (12.7 oz.).

REQUIRED REAGENTS
Code                     Unit

        Description

Quantity/test

HI  93719A-0

Hardness

Ca and Mg Reagent

0.5 mL

HI  93719B-0

Hardness

Alkaly Solution

0.5 mL

HI  93719C-0

Hardness

EDTA Solution

1 drop

HI  93719D-0

Hardness

EGTA Solution

1 drop

HI  93741-0

Iron LR

TPTZ Reagent

1 packet

Precision 

is how closely repeated measurements

agree with each other. Precision is usually expressed
as standard deviation (SD).

Accuracy

 is defined as the nearness of a test result

to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure explains
these definitions. In a laboratory using a standard
solution of 1.00 mg/L chlorine and a representative
lot of reagent, an operator obtained with a single
instrument a standard deviation of 0.02 mg/L.

Summary of Contents for HI 96741C

Page 1: ...1 w w w h a n n a i n s t c o m Instruction Manual HI 96741C Total Hardness and Iron LR ISM...

Page 2: ...LP 23 BATTERY MANAGEMENT 24 BATTERY REPLACEMENT 24 ACCESSORIES 25 WARRANTY 26 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the c...

Page 3: ...ive positive locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell It is designed to fit a cuvette with a larger neck making it easier to...

Page 4: ...4 70 mg L Iron LR 0 00 to 1 60 mg L Resolution 0 01 mg L Total Hardness 0 01 mg L Iron LR Accuracy Mg Hardness 0 11 mg L 5 of reading 25 C Ca Hardness 0 11 mg L 5 of reading 25 C Iron LR 0 01 mg L 8...

Page 5: ...oid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvette walls with hands Furthermore in order to maintain the same conditions during...

Page 6: ...t contain any debris Inverted cuvettes The sample and the zero cuvette are inverted Zero A zero reading was not taken Follow the instructions of the measurement procedure for zeroing the meter b on sa...

Page 7: ...tes that the battery is dead and must be replaced Once this indication is displayed normal operation of the instrument will be interrupted Change the battery and restart the meter Under range A blinki...

Page 8: ...s This would corrupt the reading Each time the cuvette is used the cap must be tightened to the same degree Whenever the cuvette is placed into the measurement cell it must be dry outside and complete...

Page 9: ...seconds Fill a cuvette with 10 mL of the reacted sample up to the mark and replace the cap To compensate the meter for the sample turbidity or color the measurement takes place in two phases First th...

Page 10: ...ve 4 0 mg L Nitrite ion above 0 8 mg L Chromium3 above 0 25 mg L Cobalt above 0 05 mg L Cyanide above 2 8 mg L Mercury above 0 4 mg L Nickel above 1 0 mg L Sample pH should be between 3 and 4 to avoid...

Page 11: ...ate input CALIBRATION PROCEDURE Note It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON OFF keys Warning Do not calibrate the instrument with standard solutions...

Page 12: ...RY CALIBRATION RESTORE In GLP mode the last user calibration date can be verified and the factory calibration can be restored LAST CALIBRATION DATE GLP Press RANGE GLP to edit the desired year 2009 20...

Page 13: ...ue is displayed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The rema...

Page 14: ...intained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If se...

Page 15: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mailtech hannainst com www hannainst com...

Reviews: