Instruction Manual

HI 96738C

Chlorine dioxide ISM

[email protected]

www.ponpe.com

Summary of Contents for HI 96738C

Page 1: ...1 w w w h a n n a i n s t c o m Instruction Manual HI 96738C Chlorine dioxide ISM sale ponpe com www ponpe com...

Page 2: ...information do not hesitate to e mail us at tech hannainst com All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instrume...

Page 3: ...lorine dioxide content in water samples in the 0 00 to 2 00 mg L ppm range The method is an adaptation of the Chlorophenol Red method The reagents are liquid and powder form and are supplied in bottle...

Page 4: ...lamp combined with a narrow band interference filter to guarantee both high performance and reliable results HI 96738 block diagram optical layout A microprocessor controlled special tungsten lamp em...

Page 5: ...ore factory calibration DISPLAY ELEMENTS DESCRIPTION 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The bat...

Page 6: ...ght than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over ran...

Page 7: ...general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F In order to maximize accuracy prior to a measurement follow the validation procedure to be sure that the instrument is...

Page 8: ...OCEDURE Use the validation procedure to ensure that the instrument is properly calibrated Warning Do not validate the instrument with any standard solutions other than the HANNA CAL CHECKTM Standards...

Page 9: ...orted in the CAL CHECKTM Standard Certificate If the value is found out of the specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat validation If results are s...

Page 10: ...splay will show 0 0 The meter is now zeroed and ready for calibration The blinking READ asks for reading calibration standard Remove the cuvette Place the CAL CHECKTM Standard HI 96738 11 Cuvette B in...

Page 11: ...ith three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Empty battery icon blinking if the capacity is under 10 If the battery is empty and accurate measure...

Page 12: ...harges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs pr...

Page 13: ...ents Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mailtech hannainst com www hannainst com Local Sales a...