KEYPAD DESCRIPTION
• ON/OFF: to turn the meter on and off.
• ZERO: to zero the meter prior to measurement.
• READ/TIMER: this is a bi-functional key, just press for making a measurement, or hold the key

for 3 seconds to start a pre-programmed countdown prior to measurement.

• CAL CHECK: this bi-functional key allows to perform validation of the instrument just pressing

the CAL CHECK key, or hold the key for 3 seconds to enter the 

calibration mode

.

OPERATING MODES

Measurement mode:

 default operation mode, enables both 

validation 

and 

measurement

.

Calibration mode: 

may be entered

 

by keeping CAL CHECK pressed for three seconds (the “CAL”

icon appears), it enables 

calibration 

of the instrument.

DISPLAY ELEMENTS DESCRIPTION

1) four digit main display area.
2) CAL: appears when the instrument operates in the 

calibration mode.

3) the Month and Day icons appear above the number of month and day when a date is

displayed.

4) the blinking “Battery” icon appears when the battery voltage is getting low.

13

MEASUREMENT PROCEDURE

• Turn the meter on by pressing ON/OFF.
• When the LCD displays “761”, it is ready.

• Fill the cuvet with 10 mL of unreacted sample,

up to the mark, and replace the cap.

• Place the cuvet into the holder and ensure that

the notch on the cap is positioned securely into

the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show

“ -0.0- ”. The meter is now zeroed and ready for
measurement.

• Remove the cuvet.

• Add the content of one packet of HI 95761-0

Total Chlorine Low Range reagent. Replace the
cap and shake gently for 20 seconds.

• Replace the cuvet into the holder and ensure

that the notch on the cap is positioned securely
into the groove.

Turn the meter on by pressing ON/OFF and follow instruction manual to perform either the
MEASUREMENT, VALIDATION or CALIBRATION PROCEDURE.

MEASUREMENT PROCEDURE

Summary of Contents for HI 95761C

Page 1: ...I 95761C Total Chlorine Meter for the analysis of trace Total Chlorine concentrations w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95761R2 05 06 w w w h...

Page 2: ...ake all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the ins...

Page 3: ...ACCESSORIES HI 95761 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cu...

Page 4: ...ter measures the total chlorine Cl2 content in water samples in the 0 000 to 0 500 mg L ppm range The method is an adaption of the USEPA METHOD 330 5 for wastewater The reagent is in powder form and i...

Page 5: ...sing a standard solution of 0 200 mg L chlorine and a representative lot of reagent an operator obtained with a single instrument a standard deviation of 0 004 mg L 16 CALIBRATION PROCEDURE Turn the m...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display LC...

Page 8: ...Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn the met...

Page 9: ...ibration is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the s...

Page 10: ...the right sample cuvet is inserted Inverted the sample and the zero cuvet are inverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 11 U...

Reviews: