Hanna Instruments HI 95761C Instruction Manual Download Page 4

4

The 

HI 95761 

is an auto diagnostic portable microprocessor meter that benefits from Hanna’s

years of experience as a manufacturer of analytical instruments. It has the advanced optical
system based on a special tungsten lamp and a narrow band interference filter that allows most
accurate and repeatable readings. All instruments are factory calibrated and the electronic and

optical design minimizes the need of frequent calibration.
With the powerful CAL CHECK™ 

validation

 function you are able to validate good performance of

your instrument at any time. The validation procedure is extremely user friendly. Just use the

exclusive Hanna ready-made, NIST traceable standards to verify the performance of the instrument
and recalibrate if necessary.
All instruments are splash waterproof and the lamp and filter units are protected from dust or dirt
by a transparent cup. This makes the instruments fulfill field applications. Display codes aid the

user in routine operation. The meter has an auto-shut off feature that will turn off the instrument
after 10 minutes of non use or after 1 hour if left in the 

calibration mode

.

The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same
position every time it is placed into the measurement cell. It is designed to fit a cuvet with a larger

neck making it easier to add both sample and reagents. The cuvet is made from special optical
glass to obtain best results.
The 

HI 95761

 portable meter measures the total chlorine (Cl

2

) content in water samples in the

0.000 to 0.500 mg/L (ppm) range. The method is an adaption of the USEPA METHOD 330.5 for
wastewater.
The reagent is in

 

powder form and is supplied in packets. The amount of reagent is precisely dosed

to ensure the maximum repeatability.

GENERAL DESCRIPTION

degree Celsius
US Environmental Protection Agency
degree Fahrenheit
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter

°C

:

EPA

:

°F

:

mg/L

:

mL

:

ABBREVIATIONS

17

or

• Press READ/TIMER and “SIP” will blink during

measurement.

• The instrument will show for three seconds the

CAL CHECK™ standard value.
Then the date of last calibration (e.g.: “01.08”)
appears on the display, or “01.01” if the factory
calibration was selected. In both cases the
number of the month is blinking, ready for date

input.

Note

: if the display shows “ERR” the calibration

procedure failed. Verify that both CAL CHECK™

Standards HI 95761-11 Cuvets, 

A

 and

 B

, are

free from fingerprints or dirt and that they are
inserted correctly.

DATE INPUT

• Keep READ/TIMER pressed to scroll to the desired

month number (01-12).

• When the correct month has been set, press

ZERO to confirm. Now the display will show the
day number blinking. Keep READ/TIMER pressed
to scroll to the desired day number (01-31).

Note

: it is possible to change from 

day 

to

month 

and vice versa by pressing ZERO.

• When both the day and month have been

selected, hold CAL CHECK for three seconds to
store date and calibration values. The instrument

will show for one second “Stor”, to confirm that
the new calibration data has been accepted.

• The instrument will return automatically to the

measurement mode by  displaying "761" on
the LCD.

Summary of Contents for HI 95761C

Page 1: ...I 95761C Total Chlorine Meter for the analysis of trace Total Chlorine concentrations w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95761R2 05 06 w w w h...

Page 2: ...ake all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the ins...

Page 3: ...ACCESSORIES HI 95761 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cu...

Page 4: ...ter measures the total chlorine Cl2 content in water samples in the 0 000 to 0 500 mg L ppm range The method is an adaption of the USEPA METHOD 330 5 for wastewater The reagent is in powder form and i...

Page 5: ...sing a standard solution of 0 200 mg L chlorine and a representative lot of reagent an operator obtained with a single instrument a standard deviation of 0 004 mg L 16 CALIBRATION PROCEDURE Turn the m...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display LC...

Page 8: ...Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn the met...

Page 9: ...ibration is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the s...

Page 10: ...the right sample cuvet is inserted Inverted the sample and the zero cuvet are inverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 11 U...

Reviews: