background image

9

This indicates that the meter is in a zeroed state and measurement
can be performed.

Light over range: the cuvet is not inserted correctly and an excess
ambient light is reaching the detector. If the cover is properly
installed, then contact your dealer or the nearest Hanna Customer
Service Center.

The blinking "BAT" indicates that the battery voltage is getting low
and the battery needs to be replaced.

This indicates that the battery is dead and must be replaced. Once
this indication is displayed, the meter will lock up. Change the
battery and restart.

“Configuration”: this indicates that the meter has lost its
configuration. Contact your dealer or the nearest Hanna Customer
Service Center.

CALIBRATION MODE  MESSAGES

The date of the last calibration appears on the LCD each time the

calibration mode

 is entered. If calibration is performed for the first

time, “F.CAL” appears.

This indicates that the factory calibration is selected.

The flashing digits indicate that:

1) the month can be set

2) the day can be set

“Storage”: this prompt appears for 1 second at the end of the

calibration procedure

, to indicate that the calibration data has

been stored.

“Error”: the concentration of the calibration solution used is not
correct. Repeat the 

calibration procedure

 with the right standard

solution, and verify it is not expired. If the calibration procedure fails
again, contact your dealer or the nearest Hanna Customer Service
Center.

12

MEASUREMENT PROCEDURE

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “729”, it is ready.

• Fill one cuvet with 10 mL of deionized water.

For most accurate results, use of a class A
laboratory pipette is strongly recommended.
Alternatively, fill the cuvet up to the 10 mL
mark.

• Add 2 mL of HI 93729-0 Fluoride Reagent to

the cuvet by use of the 2000 

µ

L automatic

pipette.
For a correct use of the automatic pipette,
please follow the related Instruction Sheet.

Turn the meter on by pressing ON/OFF and follow instruction manual to perform either the
MEASUREMENT, VALIDATION or CALIBRATION PROCEDURE.

2

m

L

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvet for zeroing and measurement

use the same cuvet for zeroing and measurement

use the same cuvet for zeroing and measurement

use the same cuvet for zeroing and measurement

use the same cuvet for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the 

validation procedure

 to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

MEASUREMENT PROCEDURE

Summary of Contents for HI 95729C

Page 1: ...w w w h a n n a i n s t c o m This Instrument is in Compliance with CE Directives Instruction Manual HI 95729C Fluoride Low Range ISM MAN95729CR2 10 04 w w w h a n n a i n s t c o m ...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the instrument in microwave oven For yours and the instrument safety do not use or store the instrument in hazardous environments HI 95729 is warranted for two years against defects in workmanship and mater...

Page 3: ...lorine Remover HI 95729 11 CAL CHECKTM Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs HI 731335 Caps for cuvets 4 pcs HI 731342 2000 µL automatic pipette HI 731352 tips for 2000 µL automatic pipette 4 pcs HI 93703 50 Cuvets cleaning solution 230 mL Battery replacement mu...

Page 4: ...rs The method is an adaptation of the EPA method 340 1 and Standard Methods for the Examination of Water and Wastewater 20th edition SPADNS method The reagent is in liquid form and is supplied in bottles The amount of reagent is precisely dosed by use of the supplied automatic pipette to ensure the maximum repeatability degree Celsius degree Fahrenheit Environmental Protection Agency USA Liquid Cr...

Page 5: ... lot of reagents an operator obtained with a single instrument a standard deviation of 0 03 mg L SPECIFICATIONS PRECISION AND ACCURACY 16 or Note it is possible to interrupt calibration procedure at any time by pressing ON OFF Warning do not validate or calibrate the instrument with standard solutions other than the Hanna CAL CHECKTM Standards otherwise erroneous results will be obtained For accur...

Page 6: ...w band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Use the validation procedure to ensure that the instrument is properly calibrated Warning do not validate or calibrate the instrument with standard solutions other than the Hanna CAL CHECKTM Standards otherwise erroneous results will be obtained Fo...

Page 7: ...not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display LCD 2 Cuvet Holder 3 Cuvet alignment indicator 4 ON OFF key 5 ZERO key 6 READ TIMER key 7 CAL CHECK key FUNCTIONAL DESCRIPTION 14 INTERFERENCES Chlorine interferes at all lev...

Page 8: ...in a ready state and zeroing can be performed Sampling in Progress This flashing prompt appears each time the meter is performing a measurement GUIDE TO DISPLAY CODES 13 b l a n k Replace the cap to the cuvet and shake gently a couple of times Place the cuvet with the reacted deionized water blank into the holder and ensure that the notch on the cap is positioned securely into the groove Hold READ...

Page 9: ... nearest Hanna Customer Service Center 12 MEASUREMENT PROCEDURE Turn the meter on by pressing ON OFF When the LCD displays 729 it is ready Fill one cuvet with 10 mL of deionized water For most accurate results use of a class A laboratory pipette is strongly recommended Alternatively fill the cuvet up to the 10 mL mark Add 2 mL of HI 93729 0 Fluoride Reagent to the cuvet by use of the 2000 µL autom...

Page 10: ...nteed Repeat the measurement procedure 11 The instructions listed below should be carefully followed during testing to ensure best accuracy class A pipette For adding the exact amount of sample and deionized water to the cuvet it is strongly recommended to use two class A laboratory pipettes Alternatively fill the cuvet up to the 10 mL mark the liquid in the cuvet forms a convexity on the top the ...

Reviews: