The 

HI 95729 

is an auto diagnostic portable microprocessor meter that benefits from Hanna’s

years of experience as a manufacturer of analytical instruments. It has the advanced optical
system based on a special tungsten lamp and a narrow band interference filter that allows most
accurate and repeatable readings. All instruments are factory calibrated and the electronic and
optical design minimizes the need of frequent calibration.
With the powerful CAL CHECK

TM

 

validation

 function you are able to validate good performance

of your instrument at any time. The validation procedure is extremely user friendly. Just use the
exclusive Hanna ready-made, NIST traceable standards to verify the performance of the
instrument and recalibrate if necessary.
All instruments are splash proof and the lamp and filter units are protected from dust or dirt by
a transparent cup. This makes the instruments fulfill field applications. Display codes aid the
user in routine operation. The meter has an auto-shut off feature that will turn off the instrument
after 10 minutes of non use or after 1 hour if left in the 

calibration mode

.

The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same
position every time it is placed into the measurement cell. It is designed to fit a cuvet with a
larger neck making it easier to add both sample and reagents. The cuvet is made from special
optical glass to obtain best results.
The 

HI 95729

 meter measures the Fluoride

Fluoride

Fluoride

Fluoride

Fluoride content in the 0.00 to 2.00 mg/L (ppm) range,

in drinking, surface and waste waters. The method is an adaptation of the 

EPA 

method 340.1

and

 Standard Methods for the Examination of Water and Wastewater, 20

th

 edition

, SPADNS

method.
The reagent is in

 

liquid form and is supplied in bottles. The amount of reagent is precisely dosed

by use of the supplied automatic pipette to ensure the maximum repeatability.

degree Celsius
degree Fahrenheit
Environmental Protection Agency (USA)
Liquid Crystal Display
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter
Sodium 2-(parasulfophenylazo)-1,8-dihydroxy-3,6-naphthalene disulfonate

°C

:

°F

:

EPA:

LCD:

mg/L

:

mL

:

SPADNS:

GENERAL DESCRIPTION

ABBREVIATIONS

17

or

• Press READ/TIMER and “SIP” will blink during

measurement.

• The instrument will show for three seconds the

CAL CHECK standard value.
Then the date of last calibration (e.g.:
“01.08”) appears on the display, or “01.01”
if the factory calibration was selected. In both
cases the number of the month is blinking,
ready for date input.

Note

: if the display shows “ERR” the calibration

procedure failed. Verify that the CAL CHECK

TM

Standards HI 95729-11 Cuvets, 

A

 and

 B

, are

free from fingerprints or dirt and that they are
inserted correctly.

DATE INPUT

• Keep READ/TIMER pressed to scroll to the

desired month number (01-12).

• When the correct month has been set, press

ZERO to confirm. Now the display will show
the day number blinking. Keep READ/TIMER
pressed to scroll to the desired day number
(01-31).

Note

: it is possible to change from 

day 

to

month 

and vice versa by pressing ZERO.

• When both the day and month have been

selected, hold CAL CHECK 

for three seconds

 to

store date and calibration values. The instrument
will show for one second “Stor”, to confirm
that the new calibration data has been accepted.

• The instrument will return automatically to the

measurement mode by  displaying "729" on
the LCD.

Summary of Contents for HI 95729C

Page 1: ...w w w h a n n a i n s t c o m This Instrument is in Compliance with CE Directives Instruction Manual HI 95729C Fluoride Low Range ISM MAN95729CR2 10 04 w w w h a n n a i n s t c o m ...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the instrument in microwave oven For yours and the instrument safety do not use or store the instrument in hazardous environments HI 95729 is warranted for two years against defects in workmanship and mater...

Page 3: ...lorine Remover HI 95729 11 CAL CHECKTM Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs HI 731335 Caps for cuvets 4 pcs HI 731342 2000 µL automatic pipette HI 731352 tips for 2000 µL automatic pipette 4 pcs HI 93703 50 Cuvets cleaning solution 230 mL Battery replacement mu...

Page 4: ...rs The method is an adaptation of the EPA method 340 1 and Standard Methods for the Examination of Water and Wastewater 20th edition SPADNS method The reagent is in liquid form and is supplied in bottles The amount of reagent is precisely dosed by use of the supplied automatic pipette to ensure the maximum repeatability degree Celsius degree Fahrenheit Environmental Protection Agency USA Liquid Cr...

Page 5: ... lot of reagents an operator obtained with a single instrument a standard deviation of 0 03 mg L SPECIFICATIONS PRECISION AND ACCURACY 16 or Note it is possible to interrupt calibration procedure at any time by pressing ON OFF Warning do not validate or calibrate the instrument with standard solutions other than the Hanna CAL CHECKTM Standards otherwise erroneous results will be obtained For accur...

Page 6: ...w band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Use the validation procedure to ensure that the instrument is properly calibrated Warning do not validate or calibrate the instrument with standard solutions other than the Hanna CAL CHECKTM Standards otherwise erroneous results will be obtained Fo...

Page 7: ...not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal Display LCD 2 Cuvet Holder 3 Cuvet alignment indicator 4 ON OFF key 5 ZERO key 6 READ TIMER key 7 CAL CHECK key FUNCTIONAL DESCRIPTION 14 INTERFERENCES Chlorine interferes at all lev...

Page 8: ...in a ready state and zeroing can be performed Sampling in Progress This flashing prompt appears each time the meter is performing a measurement GUIDE TO DISPLAY CODES 13 b l a n k Replace the cap to the cuvet and shake gently a couple of times Place the cuvet with the reacted deionized water blank into the holder and ensure that the notch on the cap is positioned securely into the groove Hold READ...

Page 9: ... nearest Hanna Customer Service Center 12 MEASUREMENT PROCEDURE Turn the meter on by pressing ON OFF When the LCD displays 729 it is ready Fill one cuvet with 10 mL of deionized water For most accurate results use of a class A laboratory pipette is strongly recommended Alternatively fill the cuvet up to the 10 mL mark Add 2 mL of HI 93729 0 Fluoride Reagent to the cuvet by use of the 2000 µL autom...

Page 10: ...nteed Repeat the measurement procedure 11 The instructions listed below should be carefully followed during testing to ensure best accuracy class A pipette For adding the exact amount of sample and deionized water to the cuvet it is strongly recommended to use two class A laboratory pipettes Alternatively fill the cuvet up to the 10 mL mark the liquid in the cuvet forms a convexity on the top the ...

Reviews: