background image

5

Range

0.0 to 30.0 mg/L

Resolution

0.1 mg/L

Precision

±0.5 mg/L @ 12.0 mg/L

Typical EMC Deviation

±0.2 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Light Detector

Silicon Photocell

Method

Amino Acid Method, adapted from 

Standard Methods for the

Examination of Water and Wastewater

. The reaction between

phosphate and the reagents causes a blue tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in 

measurement mode

;

after 1 hour of non-use in 

calibration mode

.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93717A-0

Phosphate HR Reagent A

10 drops

HI 93717B-0

Phosphate HR Reagent B

1 packet

Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of
12.00 mg/L phosphate and a representative lot
of reagents, an operator obtained with a single

instrument a standard deviation of 0.5 mg/L.

PRECISION AND ACCURACY

SPECIFICATIONS

16

CALIBRATION PROCEDURE

• Turn the meter on by pressing ON/OFF.
• When the LCD displays “717”, it is ready.

• Enter the 

calibration mode

 by holding CAL CHECK

for three seconds.

• The date of the last calibration appears (e.g.:

month “01”, day “08”). “F.CAL” means that the
factory calibration is selected.

Note

: at this point it is possible to reset the

instrument to restore FACTORY CALIBRATION, see

instructions page 18.

• Place the CAL CHECK™ Standard Cuvet 

A

 into

the holder and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show “-0.0-”.

The meter is now zeroed and ready for calibration.

• Remove the cuvet.

• Place the CAL CHECK™ Standard HI 95717-11

Cuvet 

B

 into the holder and ensure that the notch

on the cap is positioned securely into the groove.

Note

:

 

it is possible to interrupt calibration procedure at any time by pressing ON/OFF.

Warning

: do not validate or calibrate the instrument with standard solutions other than the Hanna CAL

CHECK™ Standards, otherwise erroneous results will be obtained.

or

CALIBRATION PROCEDURE

Summary of Contents for HI 95717C

Page 1: ...1 Instruction Manual HI 95717C Phosphate High Range ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95717R2 08 06 w w w h a n n a i n s t c o m...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Page 3: ...HI 95717 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs H...

Page 4: ...he phosphate PO4 3 content in water samples in the 0 0 to 30 0 mg L ppm range The method is an adaption from Standard Methods for the Examination of Water and Wastewater The reagents are in powder and...

Page 5: ...In a laboratory using a standard solution of 12 00 mg L phosphate and a representative lot of reagents an operator obtained with a single instrument a standard deviation of 0 5 mg L PRECISION AND ACC...

Page 6: ...mps and narrow band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OF...

Page 7: ...at the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with han...

Page 8: ...display area 2 CAL appears when the instrument operates in the calibration mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon...

Page 9: ...ation is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a c...

Page 10: ...0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zero and measurement A flashing value of the maximum conce...

Reviews: