background image

10

The flashing digits indicate that:

1) the month can be set
2) the day can be set

“Storage”: this prompt appears for 1 second at the end of the

calibration procedure

, to indicate that the calibration data has

been stored.

“Error”: the concentration of the calibration solution used is not correct.
Repeat the 

calibration procedure

 with the right standard solution, and

verify it is not expired. If the calibration procedure fails again, contact your
dealer or the nearest Hanna Customer Service Center.

ERROR  MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-to-
noise ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check that
the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement. Please
check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement. Please
check the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check if the
right sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in the

measurement procedure

 for zeroing the meter.

11

Under range. A blinking "0.00" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make
sure you use the same cuvet for reference (zero) and measurement.

A  flashing value of the maximum concentration indicates an over
range condition. The concentration of the sample is beyond the
programmed range: dilute the sample and re-run the test.

A flashing value lower than the maximum concentration indicates a
low signal-to-noise ratio condition. In this case accuracy of the result

is not guaranteed. Repeat the 

measurement procedure

.

The instructions listed below should be carefully followed during testing to ensure best accuracy.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

• Suspended matter in large amounts may cause

interference, therefore this should be removed
by prior filtration.

• For a correct filling of the cuvet: the liquid in

the cuvet forms a convexity on the top; the
bottom of this convexity must be at the same

level of the 10 mL mark.

• It is important that the sample does not contain

any debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the

measurement cell, it must be dry outside, and
completely free of fingerprints, oil or dirt. Wipe

it thoroughly with 

HI 731318

 (tissue for

wiping cuvets, see chapter ACCESSORIES) or a
lint-free cloth prior to insertion.

Summary of Contents for HI 95717C

Page 1: ...1 Instruction Manual HI 95717C Phosphate High Range ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95717R2 08 06 w w w h a n n a i n s t c o m...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Page 3: ...HI 95717 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs H...

Page 4: ...he phosphate PO4 3 content in water samples in the 0 0 to 30 0 mg L ppm range The method is an adaption from Standard Methods for the Examination of Water and Wastewater The reagents are in powder and...

Page 5: ...In a laboratory using a standard solution of 12 00 mg L phosphate and a representative lot of reagents an operator obtained with a single instrument a standard deviation of 0 5 mg L PRECISION AND ACC...

Page 6: ...mps and narrow band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OF...

Page 7: ...at the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with han...

Page 8: ...display area 2 CAL appears when the instrument operates in the calibration mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon...

Page 9: ...ation is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a c...

Page 10: ...0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zero and measurement A flashing value of the maximum conce...

Reviews: