background image

9

This prompt appears for 1 second each time the instrument is
turned on.

This indicates that the meter is in a ready state and zeroing can
be performed.

Sampling in Progress. This flashing prompt appears each time the
meter is performing a measurement.

This indicates that the meter is in a zeroed state and measurement
can be performed.

Light over range: the cuvet is not inserted correctly and an excess
ambient light is reaching the detector. If the cover is properly
installed, then contact your dealer or the nearest Hanna Customer
Service Center.

The blinking "BAT" indicates that the battery voltage is getting low
and the battery needs to be replaced.

This indicates that the battery is dead and must be replaced. Once
this indication is displayed, the meter will lock up. Change the
battery and restart.

“Configuration”: this indicates that the meter has lost its
configuration. Contact your dealer or the nearest Hanna Customer
Service Center.

CALIBRATION MODE  MESSAGES

The date of the last calibration appears on the LCD each time the

calibration mode

 is entered. If calibration is performed for the first

time, “F.CAL” appears.

This indicates that the factory calibration is selected.

GUIDE TO DISPLAY CODES

12

• It is important that the sample does not contain

any debris. This would corrupt the reading.

• In order to avoid reagent leaking and to obtain

more accurate measurements, it s recommended

to close the cuvet first with the supplied HDPE

plastic stopper and then with the black cap.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free

of fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 (tissue for wiping cuvets, see chapter

ACCESSORIES) or a lint-free cloth prior to insertion.

• Do not let the reacted sample stand too long after

reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row,

but it is recommended to take a new zero reading

for each sample and to use the same cuvet for

zeroing and measurement.

• After the reading it is important to discard

immediately the sample, otherwise the glass might
become permanently stained.

• All the reaction times reported in this manual are

referred to 20°C (68°F). As a general rule of

thumb, they should be doubled at 10°C (50°F)
and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a

measurement follow the 

validation procedure

to be sure that the instrument is properly

calibrated. If necessary, calibrate the instrument.

Summary of Contents for HI 95715

Page 1: ...w w w h a n n a c a n c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95715 Ammonia Medium Range ISM MAN95715R1 09 03 w w w h a n n a c a n c o m...

Page 2: ...ry steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the instrument in micr...

Page 3: ...s 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs HI 731335 Caps for cuvets 4 pcs HI 937...

Page 4: ...ater samples in the 0 00 to 9 99 mg L ppm range The method is an adaptation from ASTM Manual of Water and Environmental Technology D1426 93 Nessler Method The reagents are in liquid form and are suppl...

Page 5: ...ry using a standard solution of 6 00 mg L ammonia nitrogen NH3 N and a representative lot of reagents an operator obtained with a single instrument a standard deviation of 0 12 mg L PRECISION AND ACCU...

Page 6: ...a narrow band interference filter to guarantee both high performance and reliable results HI 95715 block diagram optical layout PRINCIPLE OF OPERATION 15 VALIDATION PROCEDURE Turn the meter on by pres...

Page 7: ...ce due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measurin...

Page 8: ...Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn the meter on by p...

Page 9: ...n is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading In order to avoid reagent leaking and to obtain more accurate measure...

Page 10: ...s not taken Follow the instruction in the measurement procedure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the proc...

Reviews: