The 

HI 95715 

is an auto diagnostic portable microprocessor meter that benefits from Hanna’s

years of experience as a manufacturer of analytical instruments. It has the advanced optical
system based on a LED light source and a narrow band interference filter that allows most
accurate and repeatable readings. All instruments are factory calibrated and the electronic and
optical design minimizes the need of frequent calibration.
With the powerful CAL CHECK 

validation

 function you are able to validate good performance of

your instrument at any time. The validation procedure is extremely user friendly. Just use the
exclusive Hanna ready-made, NIST traceable standards to verify the performance of the
instrument and recalibrate if necessary.
All instruments are splash proof and the LED (Light Emitting Diode) and filter units are protected
from dust or dirt by a transparent cup. This makes the instruments fulfill field applications.
Display codes aid the user in routine operation. The meter has an auto-shut off feature that will
turn off the instrument  after 10 minutes of non use or after 1 hour if left in the 

calibration mode

.

The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same
position every time it is placed into the measurement cell. It is designed to fit a cuvet with a larger
neck making it easier to add both sample and reagents. The cuvet is made from special optical
glass to obtain best results.
The 

HI 95715

 meter measures the ammonia nitrogen (NH

3

-N) content in water samples in the

0.00 to 9.99 mg/L (ppm) range. The method is an adaptation from 

ASTM Manual of Water and

Environmental Technology, D1426-93

, Nessler Method.

The reagents are in

 

liquid form and are supplied in bottles. The amount of reagent is precisely

dosed to ensure the maximum repeatability.

GENERAL DESCRIPTION

ABBREVIATIONS

American Society for Testing and Materials
degree Celsius
degree Fahrenheit
Liquid Crystal Display
Light Emitting Diode
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter

ASTM:

°C

:

°F

:

LCD:
LED:

mg/L

:

mL

:

17

or

• Press READ/TIMER and “SIP” will blink during

measurement.

• The instrument will show for three seconds the

Cal Check standard value.
Then the date of last calibration (e.g.: “01.08”)
appears on the display, or “01.01” if the factory
calibration was selected. In both cases the
number of the month is blinking, ready for
date input.

Note

: if the display shows “ERR” the calibration

procedure failed. Verify that the CAL CHECK
Standards HI 95715-11 Cuvets, 

A

 and

 B

, are

free from fingerprints or dirt and that they are
inserted correctly.

DATE INPUT

• Keep READ/TIMER pressed to scroll to the

desired month number (01-12).

• When the correct month has been set, press

ZERO to confirm. Now the display will show the
day number blinking. Keep READ/TIMER pressed
to scroll to the desired day number (01-31).

Note

: it is possible to change from 

day 

to

month 

and vice versa by pressing ZERO.

• When both the day and month have been

selected, hold CAL CHECK for three seconds to
store date and calibration values. The instrument
will show for one second “Stor”, to confirm
that the new calibration data has been accepted.

• The instrument will return automatically to the

measurement mode by  displaying "715" on
the LCD.

Summary of Contents for HI 95715

Page 1: ...w w w h a n n a c a n c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95715 Ammonia Medium Range ISM MAN95715R1 09 03 w w w h a n n a c a n c o m...

Page 2: ...ry steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the instrument in micr...

Page 3: ...s 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs HI 731335 Caps for cuvets 4 pcs HI 937...

Page 4: ...ater samples in the 0 00 to 9 99 mg L ppm range The method is an adaptation from ASTM Manual of Water and Environmental Technology D1426 93 Nessler Method The reagents are in liquid form and are suppl...

Page 5: ...ry using a standard solution of 6 00 mg L ammonia nitrogen NH3 N and a representative lot of reagents an operator obtained with a single instrument a standard deviation of 0 12 mg L PRECISION AND ACCU...

Page 6: ...a narrow band interference filter to guarantee both high performance and reliable results HI 95715 block diagram optical layout PRINCIPLE OF OPERATION 15 VALIDATION PROCEDURE Turn the meter on by pres...

Page 7: ...ce due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measurin...

Page 8: ...Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn the meter on by p...

Page 9: ...n is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading In order to avoid reagent leaking and to obtain more accurate measure...

Page 10: ...s not taken Follow the instruction in the measurement procedure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the proc...

Reviews: