USER CALIBRATION

The 

HI93551N

HI93551R

HI93552

 and

 HI93552R

 models can be

calibrated at 0°C by using an ice bath.
• Prepare an ice bath with approximately equal volumes of distilled

water and chopped ice made from distilled water.

• Immerse the temperature probe in the center of the ice bath,

taking care not to touch the ice with the probe tip.

• Ensure that the meter is measuring a temperature within ±3°C.
• To enter the Calibration mode, press and hold the CAL button for

about 5 seconds.

• The CAL tag turns on to indicate that

the Calibration mode has been entered.

Note

: If the measurement is outside the

±3°C window, the meter does not
enter Calibration mode.

• When the meter reaches the stability condition, which is detected

when the measurement remains within ±0.2°C for 5 seconds,
the calibration is accepted and the reading becomes 0°C (32°F).

• The meter then automatically returns to normal mode.

Note

: To exit the Calibration mode at any time, press the CAL key.

Note

: User calibration cannot be entered in relative, average  or in

T1-T2 mode.

Note

: User calibration is only performed on the current channel

displayed (T1 or T2).

FACTORY RECALIBRATION

All Hanna thermometers have been accurately pre-calibrated at the
factory.
It is generally recommended to have all thermometers recalibrated at
least once a year.
For an accurate recalibration, contact your nearest Hanna Customer
Service Center.

Summary of Contents for HI 93542

Page 1: ...Instruction Manual HI93551 HI93551N HI93551R HI93542 HI93552 HI93552R K J T Thermometers www hannainst com...

Page 2: ...42 HI93552 HI93551R ans HI93552R are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions Electrodes and...

Page 3: ...accurate high resolution measure ments in a wide temperature reange Standard features include dual level LCD which allows continuous displaying of High and Low temperature values C F selection HOLD f...

Page 4: ...0 1 C 200 0 to 999 9 C 0 1 F 149 9 to 999 9 F 0 2 F 328 0 to 150 0 F 1 F 1000 to 1832 F T 0 1 C 149 9 to 400 0 C 0 2 C 200 0 to 150 0 C 0 1 F 0 0 to 752 0 F 0 2 F 270 0 to 0 1 F 0 3 F 328 0 to 270 1...

Page 5: ...freeze the reading on display C F change reading unit C or F KJT select the thermocouple type CLR clear the HI and LO values CAL HI93551N only press and hold for about 5 seconds to enter the Calibrat...

Page 6: ...9 9 to 999 9 F 0 2 F 328 0 to 150 0 F 1 F 1000 to 1832 F T 0 1 C 149 9 to 400 0 C 0 2 C 200 0 to 150 0 C 0 1 F 0 0 to 752 0 F 0 2 F 270 0 to 0 1 F 0 3 F 328 0 to 270 1 F Accuracy 0 5 C 100 0 to 999 9...

Page 7: ...mode KJT select the thermocouple type ALT HI93552 only enable the second function keys the ALT tag turns on to indicate that the second functions are enabled Note The ALT key can be released before p...

Page 8: ...328 0 to 150 0 F 1 F 1000 to 1832 F T 0 1 C 149 9 to 400 0 C 0 2 C 200 0 to 150 0 C 0 1 F 0 0 to 752 0 F 0 2 F 270 0 to 0 1 F 0 3 F 328 0 to 270 1 F Accuracy 0 5 C 100 0 to 999 9 C 1 C outside 20 C 6...

Page 9: ...r T1 T2 measuring mode ALT enable the second function keys the ALT tag turns on to indicate that the second functions are enabled Note The ALT key can be released before pressing the second function k...

Page 10: ...ress the ON OFF key Note The meters are provided with an acoustic signal feature activated when buttons are pressed which can be disabled using a switch located in the battery compartment see figure o...

Page 11: ...W TEMPERATURES The maximum and minimum tempera tures are continuously monitored and displayed on the lower portion of the LCD Note When the reading goes over range or the probe is removed the Hi and L...

Page 12: ...l buttons wait a few seconds and the meter will return to normal measurement mode BACKLIGHT FEATURE The HI93551R HI93552 and HI93552R mod els are provided with a backlight feature which can be easily...

Page 13: ...main body of the LCD shows the current temperature while the lower portion displays the Hi Lo limits for that channel When the meter enters relative mode the REL tag lights up and the current temperat...

Page 14: ...g AVG tag The average value is lost and will not restore even if the condition restores To start the average cycle again either press CLR or re enter the average mode While in T1 T2 not for HI93551R p...

Page 15: ...fter download is complete use thesetup exe file to install the software The 32 character data string is structured as follows Main portion of the LCD byte 0 probe type K thermocouple k bytes 1 2 measu...

Page 16: ...outside the 3 C window the meter does not enter Calibration mode When the meter reaches the stability condition which is detected when the measurement remains within 0 2 C for 5 seconds the calibratio...

Page 17: ...hile paying attention to their polarity Reattach the back making sure that the gasket is in place and tighten the 4 screws to ensure a watertight seal Battery replacement must only take place in a non...

Page 18: ...mperature flexible wire probe without handle max 1100 C 2000 F HI766F1 Flexible wire probe w o handle max 480 C 900 F HI766TR1 Penetration probe max 250 C 482 F HI766TR2 Penetration long probe max 250...

Page 19: ...lectronicEquipment Theproduct should not be treated as household waste Insteadhanditovertotheappropriatecollectionpointforthe recycling of electrical and electronic equipment which will conserve natur...

Page 20: ...20 MANKJT 06 19 WorldHeadquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI02895USA www hannainst com PrintedinROMANIA...

Reviews: