VIEWING LOGGED DATA

• Press ALT+VIEW keys: the last sample will be displayed.

• Press the “right arrow” key to scroll among turbidity value, date and time of

the sample.

• Press the “up/down arrow” keys to view next sample and use the “right arrow”

to scroll the date and time.

• Press ALT+VIEW again to exit and return to measurement mode.

VIEWING AND SETTING DATE AND TIME

VIEWING

• Press  ALT+DATE to toggle between date and time.

SETTING

• Press and hold ALT+DATE  for 3-4 seconds.

• Press the “up/down arrow” keys to set the year.

• Press the “right arrow” to scroll to month & day, then use the “up/down arrow”

keys to set the date.

• Press the “right arrow” again to scroll to time, then use the “up/down arrow”

keys to set the time.

• Press the “right arrow” again to save and exit from setup mode.

Note: Press ALT+DATE to exit without saving.

• Press the READ key and the LCD will display a blinking “SIP” (Sampling In

Progress). The measured value will appear after approximately 25 seconds.

SOURCES OF INTERFERENCE

• Presence of floating debris and coarse sediment which settle out rapidly will give

false readings.

• The infrared light source used for 

HI 93124

 Haze meter can effectively minimize

errors due to coloured dissolved substances. This effect, named “true colour”,
is a common interference for most commercially available instruments operating
in the range of visible light.

• Air bubbles and the effect of vibrations that disturb the surface of the sample

will give false results.

• Dirty glassware could also affect readings along with scratched or edged cuvette.

LOG-ON-DEMAND

The meter can store up to 199 measurements in the internal memory. Press
the STO key: the last measured value will be stored and the display will show
the reference number of the stored sample.

When the maximum number of samples (199) is
reached, the LCD will display the "FULL" message
to warn the user that next value will overwrite the
oldest sample recorded.

All samples will be shifted (reference number reduced by one) and the ref. number
one will continue to indicate the oldest value. To confirm storing of the current
sample, press the STO button again.

CLEARING LOGGED DATA

For clearing all data stored in memory, press
ALT+CLR keys: the meter will ask for confirmation
by displaying the "CLr" message.

Press ALT+CLR  again to clear memory or any button to abort the operation.

To communicate with the instrument  through the 

HI92000

 software, simply connect

the 9-pin serial port of the PC to the 5-pin connector of the meter, by using the

HI920011

 cable (optional).

The meter will maintain its full functionality, being also able to perform data transfer
while operating.

Notes

:

• RS232 communication is not allowed during calibration.
• The auto-off feature is disabled when the meter is connected to a PC.

RS232 CONNECTION

Summary of Contents for HI 93124

Page 1: ...1 Instruction Manual HI 93124 Portable Logging Haze Meter w w w h a n n a i n s t c o m...

Page 2: ...instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance are not covered If service is required contact...

Page 3: ...epolyphenol polymerizes A range of stabilization treatments is available for avoiding haze problems and theproducthastobecontrolledonseveralstepsduringbrewingprocess inparticular after filtration and...

Page 4: ...withalint freetissue HI731318 before inserting into the measure ment cell Thecuvettemustbecompletelyfreeoffingerprints andotheroilordirt particularlyintheareawhere the light goes through approximately...

Page 5: ...availableinstrumentsoperating in the range of visible light Air bubbles and the effect of vibrations that disturb the surface of the sample willgivefalseresults Dirtyglasswarecouldalsoaffectreadingsal...

Page 6: ...from stock solutions TheconsistencyofHI93124readingsbyusingbothstandardshasbeenseparately established by Advanced Polymer Systems and Hanna Instruments Additional documentation about the formazine sta...

Page 7: ...ette cap always to the same degree Beforeinsertingthecuvetteintotheinstrument wipeitwithsoft lint freetissue Handlecuvettessothatnofingerprintscangetontheareaswherelightpasses approximately 2 cm 1 inc...

Page 8: ...4 1 HI 93124 2 HI731318 Tissue for wiping cuvettes 4 pcs HI 731321 Spare measurement cuvettes 4 pcs HI92000 Windows compatibleapplicationsoftware HI920011 5 to 9 pin RS232 connection cable HI 93703 50...

Page 9: ...3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiw...

Reviews: