The instrument is supplied complete with batteries. Before starting operations, install
the batteries while paying attention to their polarity (see also “Battery Replacement”
section). To maximize the battery life the meter automatically switches off after 5
minutes of non-use. For reactivating it, simply press the ON/OFF key.

TAKING MEASUREMENTS

Turn the meter ON by pressing the ON/OFF key.

At start-up the meter will show the full display for
a few seconds, followed by the indication of the
remaining battery percentage.

When the LCD displays “-- --” the meter is ready
to measure.

Fill a clean cuvette up to 0.5 cm (one quarter inch)
from its rim with the thoroughly agitated sample.

Allow sufficient time for bubbles to escape before securing the cap.

Note:

 

Do not over-tighten the cap.

• Wipe the cuvette thoroughly with a lint-free tissue

(

HI731318

) before inserting into the measure-

ment cell.
The cuvette must be completely free of fingerprints
and other oil or dirt, particularly in the area where
the light goes through (approximately the bot-
tom 2 cm/1 inch of the cuvette).

• Place the cuvette into the cell and check that the

notch on the cap is positioned securely into the
groove.

• The mark on the cuvette cap should point towards

the LCD.

OPERATIONAL GUIDE

SPECIFICATIONS

Range

0 to 10.00 EBC

10 to 250 EBC

Resolution

0.01 EBC or 1 EBC

Accuracy                      

±0.20 EBC ±3% of reading

 

(0.00 to 10.00 EBC)

±1 EBC ±5% of reading (outside)

Typical EMC Deviation

 ±1% F.S.

Light Source

Infrared LED @890 nm

Light Source Life

Life of the instrument

Light Detector

Silicon Photocell

Power Source

4 x 1.5V AA alkaline batteries

Battery Life

60 hours or 900 measurements

Auto-off

After 5 minutes of non-use

Environment

0 to 50

°

C (32 to 122

°

F);

RH max 95% non-condensing

Dimensions

220 x 82 x 66 mm (8.7 x 3.2 x 2.6")

Weight

510 g (1.1 lb.)

Summary of Contents for HI 93124

Page 1: ...1 Instruction Manual HI 93124 Portable Logging Haze Meter w w w h a n n a i n s t c o m...

Page 2: ...instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance are not covered If service is required contact...

Page 3: ...epolyphenol polymerizes A range of stabilization treatments is available for avoiding haze problems and theproducthastobecontrolledonseveralstepsduringbrewingprocess inparticular after filtration and...

Page 4: ...withalint freetissue HI731318 before inserting into the measure ment cell Thecuvettemustbecompletelyfreeoffingerprints andotheroilordirt particularlyintheareawhere the light goes through approximately...

Page 5: ...availableinstrumentsoperating in the range of visible light Air bubbles and the effect of vibrations that disturb the surface of the sample willgivefalseresults Dirtyglasswarecouldalsoaffectreadingsal...

Page 6: ...from stock solutions TheconsistencyofHI93124readingsbyusingbothstandardshasbeenseparately established by Advanced Polymer Systems and Hanna Instruments Additional documentation about the formazine sta...

Page 7: ...ette cap always to the same degree Beforeinsertingthecuvetteintotheinstrument wipeitwithsoft lint freetissue Handlecuvettessothatnofingerprintscangetontheareaswherelightpasses approximately 2 cm 1 inc...

Page 8: ...4 1 HI 93124 2 HI731318 Tissue for wiping cuvettes 4 pcs HI 731321 Spare measurement cuvettes 4 pcs HI92000 Windows compatibleapplicationsoftware HI920011 5 to 9 pin RS232 connection cable HI 93703 50...

Page 9: ...3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiw...

Reviews: