Hanna Instruments HI 93114 Instruction Manual Download Page 6

11

10

10) CAL key:

If pressed during calibration,
the calibration procedure will
be aborted and the last cali-
bration data will be retrieved.

If pressed together with the
ALT key for less then 3 sec-
onds, the diagnostic mode will
be entered.

If pressed together with the
ALT key again, the meter will
quit diagnostic mode.

If pressed for more then 3
seconds, a blinking “CAL” will
appear and the calibration
procedure is entered.

11) GLP/Abs key: In ion specific mode, it will

toggle concentration/absor-
bance readings on main LCD.

In turbidity mode, GLP data
(date, time and the 2 calibra-
tion values) will be shown.

If pressed in time/date setup
mode, the meter will quit cur-
rent mode without making any
change.

12)   key

Scrolls upwards through the
parameters to be measured.

In calibration and diagnostic
modes, increments the blink-
ing digit by one.

If pressed together with ALT
while the meter is in logging
mode, the upper LCD will
show the stored data (date,
time, value).

13) ALT key:

Activates alternative functions.

14)   key:

Scrolls downwards through the
parameters to be measured.

2) Secondary LCD:

The three-digit lower level shows the cur-
rent measurement mode (“F CL”, “t CL”, or
“tr”), and diagnostic or calibration modes,
such as “d 11”, “2 Fn”, “5 c1”.

3) DATE:

Indicates that the main LCD
is showing the current date,
the date of last calibration or
the date of logged measure-
ment in memory.

4) TIME:

Indicates that the main LCD
is showing the current time,
the time of last calibration or
the time of logged measure-
ment in memory.

5) LOBAT:

Blinks to warn the user of low
battery voltage.

6) LOG:

Blinks to indicate that the
meter is in the scroll mode
viewing the logged data.

If fixed, it indicates that the
meter is in the log mode and
every reading will be stored
in memory.

7) ON/OFF key:Turns the meter on and off.

8) ZERO /

 key: In ion specific mode, it ze-

ros the sample.

In calibration and diagnostic
modes, it works as ENTER.

In turbidity mode, it is not
used.

9) READ /

 key: Takes concentration/tur-

bidity measurement of the
sample.

In diagnostic or calibration
mode, shifts the flashing digit
to the right.

Summary of Contents for HI 93114

Page 1: ...Instruction Manual HI 93114 Turbidity Chlorine Measurements w w w h a n n a i n s t c o m...

Page 2: ...ectly Any defective item must be re turned in its original packaging with the supplied accessories 1 NTU Nephelometric Turbidity Unit 1 FTU Formazine Turbidity Unit DearCustomer ThankyouforchoosingaHA...

Page 3: ...PA compliant turbidity mea surements are performed quickly and reliably withthislab grademicroprocessor basedfield instrument HI93114isparticularlyusefulforpotablewater measurementsandmunicipalities W...

Page 4: ...nce determined by the emitted radiation I An LED Light Emitting Diode emits radiation at a relatively narrow spectrum supplying the system with the intensity Io LED EMITTED LIGHT CUVET MICROPROCESSOR...

Page 5: ...the radiation I that is not absorbed by the sample and converts it into an electric current The microprocessor converts the value into the desired measuring unit and displays it on the LCD The measure...

Page 6: ...ored data date time value 13 ALT key Activates alternative functions 14 key Scrolls downwards through the parameters to be measured 2 Secondary LCD The three digit lower level shows the cur rent measu...

Page 7: ...ur minute data will be stored in memory SPECIFICATIONS 1 NTU 1 FTU In calibration and diagnostic modes decreases the blink ing digit by one If pressed together with ALT while the meter is in logging m...

Page 8: ...e notch on the cap is positioned securely into the groove Press ZERO CURRENT TIME DATE RECALL To recall current TIME DATE press and hold the ALT and READ keys together The cur rent time and an intermi...

Page 9: ...e been stored Add the appropriate reagent into the blank sample cuvet Shake and allow a few sec onds for color to develop For the Free or Total Chlorine add the con tent of its respective packet Repla...

Page 10: ...Turn the meter on and press both ALT and CAL momentarily The LCD will show four dashes on the upper and d 00 on the lower part of the LCD with the second 0 blinking Using the and keys change the lowe...

Page 11: ...vet holder and make sure that the notch is positioned securely into the groove Press the key The display will indicate SIP again The display will blink Lt for several sec onds indicating that the LED...

Page 12: ...point customized calibration To enter the colorimetric calibration mode press momentarily both the ALT and CAL keys The LCD will show four dashes and d00 After the second calibration point is memo riz...

Page 13: ...ion point is memo rized the unit will store the calibration data time and date in the EEPROM while intermit tently indicating CAL and Stor for several seconds Using the and keys set the lower part of...

Page 14: ...brates span in turbidimetric mode 31 Calibrates span in colorimetric mode 40 Clears the logged memory The following diagnostic modes are reserved for authorized service technicians 00 Shows the Blank...

Page 15: ...r can easily turn the log ging mode on and off review the logged memory review the current lot number and clean the buffer memory HI 93114 also reminds the user if its memory is full TURNING THE LOG M...

Page 16: ...ngs in its non volatile memory and the display will flash CAL and Stor alternately for several sec onds USER SELECTABLE SHUTDOWN With HI 93114 the users can customize the shutdown time to save power T...

Page 17: ...er ies as soon as possible BA Exhausted batteries change all bat teries immediately LO Low light level is received during the zeroing procedure check the cuvet for scratches and ensure that the sample...

Page 18: ...according to instructions This warranty is limited to repair or replace ment free of charge Damages due to acci dents misuse tampering or lack of prescribed maintenance are not covered If service is r...

Page 19: ...3 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 27...

Reviews: