Hanna Instruments HI 93114 Instruction Manual Download Page 15

29

28

of the display to remind the user that every time
a measurement is taken, the value is stored in
the next available lot number.

REVIEWING THE LOGGED BUFFER

To review the memo-
rized values, press to-

gether ALT and  .

The meter will scroll all the data in the buffer
showing the lot number, value, date and time.

e.g. The first recorded reading in the buffer is

lot # 0, 0.35 mg/L of Free Chlorine,
memorized on 23

rd

 August at 3:34 pm;

The second logged data relates to lot
# 1, 1.35 mg/L of the customized param-
eter, logged on 23

rd

 August at 3:55 pm.

First lot

Second lot

CLEARING THE MEMORY

After all the buffer (memory) is taken up, the
LCD will blink “Full”.

LOGGING

HI 93114 allows user to log 25 time/day-tagged
measurements. User can easily turn the log-
ging mode on and off, review the logged
memory, review the current lot number and
clean the buffer (memory). HI 93114 also
reminds the user if its memory is full.

TURNING THE LOG MODE ON OR OFF

Enter the diagnostic
mode by pressing ALT
and CAL together.

Select mode 12 and press the

 key.

The display will show the current (vacant) lot
together with “LOG” if the log mode is on.
Otherwise it will show “----” if the log mode is
off.

By pressing the 

 key, the meter toggles

between the Log on and off positions. If the
log on mode is selected, every time a mea-
surement is taken (READ pressed) the rel-
evant values will be stored in the current
(vacant) lot number.

To quit diagnostic
mode, press ALT and
CAL together again.

The LCD will then show “CAL” and “Stor” for a
few seconds. If the log on mode was selected
“LOG” will appear on the bottom left hand side

+

To quit diagnostic mode,
press the ALT and CAL
keys together again.

+

+

+

Summary of Contents for HI 93114

Page 1: ...Instruction Manual HI 93114 Turbidity Chlorine Measurements w w w h a n n a i n s t c o m...

Page 2: ...ectly Any defective item must be re turned in its original packaging with the supplied accessories 1 NTU Nephelometric Turbidity Unit 1 FTU Formazine Turbidity Unit DearCustomer ThankyouforchoosingaHA...

Page 3: ...PA compliant turbidity mea surements are performed quickly and reliably withthislab grademicroprocessor basedfield instrument HI93114isparticularlyusefulforpotablewater measurementsandmunicipalities W...

Page 4: ...nce determined by the emitted radiation I An LED Light Emitting Diode emits radiation at a relatively narrow spectrum supplying the system with the intensity Io LED EMITTED LIGHT CUVET MICROPROCESSOR...

Page 5: ...the radiation I that is not absorbed by the sample and converts it into an electric current The microprocessor converts the value into the desired measuring unit and displays it on the LCD The measure...

Page 6: ...ored data date time value 13 ALT key Activates alternative functions 14 key Scrolls downwards through the parameters to be measured 2 Secondary LCD The three digit lower level shows the cur rent measu...

Page 7: ...ur minute data will be stored in memory SPECIFICATIONS 1 NTU 1 FTU In calibration and diagnostic modes decreases the blink ing digit by one If pressed together with ALT while the meter is in logging m...

Page 8: ...e notch on the cap is positioned securely into the groove Press ZERO CURRENT TIME DATE RECALL To recall current TIME DATE press and hold the ALT and READ keys together The cur rent time and an intermi...

Page 9: ...e been stored Add the appropriate reagent into the blank sample cuvet Shake and allow a few sec onds for color to develop For the Free or Total Chlorine add the con tent of its respective packet Repla...

Page 10: ...Turn the meter on and press both ALT and CAL momentarily The LCD will show four dashes on the upper and d 00 on the lower part of the LCD with the second 0 blinking Using the and keys change the lowe...

Page 11: ...vet holder and make sure that the notch is positioned securely into the groove Press the key The display will indicate SIP again The display will blink Lt for several sec onds indicating that the LED...

Page 12: ...point customized calibration To enter the colorimetric calibration mode press momentarily both the ALT and CAL keys The LCD will show four dashes and d00 After the second calibration point is memo riz...

Page 13: ...ion point is memo rized the unit will store the calibration data time and date in the EEPROM while intermit tently indicating CAL and Stor for several seconds Using the and keys set the lower part of...

Page 14: ...brates span in turbidimetric mode 31 Calibrates span in colorimetric mode 40 Clears the logged memory The following diagnostic modes are reserved for authorized service technicians 00 Shows the Blank...

Page 15: ...r can easily turn the log ging mode on and off review the logged memory review the current lot number and clean the buffer memory HI 93114 also reminds the user if its memory is full TURNING THE LOG M...

Page 16: ...ngs in its non volatile memory and the display will flash CAL and Stor alternately for several sec onds USER SELECTABLE SHUTDOWN With HI 93114 the users can customize the shutdown time to save power T...

Page 17: ...er ies as soon as possible BA Exhausted batteries change all bat teries immediately LO Low light level is received during the zeroing procedure check the cuvet for scratches and ensure that the sample...

Page 18: ...according to instructions This warranty is limited to repair or replace ment free of charge Damages due to acci dents misuse tampering or lack of prescribed maintenance are not covered If service is r...

Page 19: ...3 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 27...

Reviews: