8

INSTRUCTION MANUAL

1.

Install the Burette Dispensing Tube

The dispensing tube is mounted on the right side of the burette. This tube needs to be
replaced with a longer one in order to reach the Autosampler.
To remove the dispensing tube follow these steps:

• Slide up the tube protector (3).

• Remove the tube lock (4) from the burette holder.

• Unscrew the fitting (2).

• Remove the tube.

To attach the new dispensing tube follow these steps:

• Insert the flat-shaped end of the dispensing tube into the valve outlet, and screw in

the fitting so that the highest of its 9 cuts stays vertically in the final position (2).

• Bend the tube up into the vertical position to enter the highest cut of the fitting (5).

• Put on the tube lock on the fitting (4).

• Slide down tightly the tube protector (3) into the dedicated gap of the tube lock.

3

1

4

1

2

1

2

5

1

4

5

1

3

1

SETUP

Summary of Contents for HI 921

Page 1: ...1 INSTRUCTION MANUAL HI 921 AUTOSAMPLER Revision 1 1 www hannainst com ...

Page 2: ...2 INSTRUCTION MANUAL ...

Page 3: ...r This manual will provide you with the necessary information to correctly set up the autosampler For information on the autosampler functinality see Chapter 12 in the HI 902 Potentiometric Titrator Manual 2015 Hanna Instruments All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc 584 Park East Drive Woonsoc...

Page 4: ...AM 7 SETUP 8 1 Install the Burette Dispensing Tube 8 2 Attach the Cable Carrier 9 3 Set Up the Dispenser Head 10 4 Connect the Pump Tubes 11 5 Install the Tray 12 6 Electrical Connections 13 OPERATION 14 MAINTENANCE 15 Upgrading Firmware 17 TECHNICAL SPECIFICATIONS 18 AUTOSAMPLER COMPONENTS 19 ...

Page 5: ...ence The autosampler and the accessories are shipped in a single box containing ITEM QUANTITY 1 Autosampler 1 pc 2 Control Panel 1 pc 3 Tray Locking Screw 1 pc 4 Titrant dispensing tube 1 pc 5 Electrode Holder 1 pc 6 Communication cable 1 pc 7 BNC cable 1 pc 8 Temperature Sensor 1 pc 9 USB Memory Stick HI 902 Software Update included 1 pc 10 Quick Start 1 pc 11 Quality Certificate 1 pc See Autosam...

Page 6: ...working conditions Severe vibrations Direct sunlight Atmospheric relative humidity above 95 non condensing Environment temperatures below 10 C and above 40 C Explosion hazards 6 Have the Autosampler serviced by qualified service personnel only 7 Keep hands and objects clear of moving parts during operation In an emergency situation the Autosampler can be stopped immediately by pressing either of t...

Page 7: ...7 INSTRUCTION MANUAL INSTALLATION DIAGRAM Rear View Front View Side View ...

Page 8: ...e burette holder Unscrew the fitting 2 Remove the tube To attach the new dispensing tube follow these steps Insert the flat shaped end of the dispensing tube into the valve outlet and screw in the fitting so that the highest of its 9 cuts stays vertically in the final position 2 Bend the tube up into the vertical position to enter the highest cut of the fitting 5 Put on the tube lock on the fittin...

Page 9: ...RUCTION MANUAL 2 Attach the Cable Carrier To attach the cable carrier remove the three 3 screws from the top rear panel A Align the screw holes on the cable carrier B to the rear panel and replace the screws ...

Page 10: ...ure sensor stirrer and aspiration tube optional into the dedicated holes in the electrode holder Push them down until they are in a stable position Insert burette and auxiliary tubing into the appropriate holders Close clips on cable carrier once all cables and tubing are run ...

Page 11: ...11 INSTRUCTION MANUAL 4 Connect the Pump Tubes ...

Page 12: ...e the tray on the turntable with Beaker 1 under the dispenser Make sure the turntable mounting pins are aligned with the tray Lock the tray in place with the locking screw Do not over tighten the locking screw 5 Install the Tray ...

Page 13: ...trator to the 5 pin threaded connector Connect the control panel to the 4 pin threaded connector Connect the overhead stirrer optional to the 4 pin mini DIN connector Connect a USB barcode reader optional to the USB slot 6 Electrical Connections ...

Page 14: ...used waiting for user action Red flashing Error Red steady Firmware corrupted Autosampler initializing first 3 seconds after power on Control Panel The control panel displays the current status of the Autosampler and can be used to manually control the Autosampler while it is idle green steady status lights Move the dispenser head up down Rotate the turnable clockwise counter clockwise Power the a...

Page 15: ... cover off of the pump 1 Remove the plastic rotor and tubing 2 Thoroughly grease the outside of the replacement tube and the rollers of the rotor Wrap the replacement tubing around the rotor and press into the pump housing 3 The tubing in contact with the rollers will need to be compressed with a finger in order to fit in the housing Replace the clear plastic cover MAINTENANCE ...

Page 16: ...rcuit board 5 Pull the pump out of the Autosampler tower If the pump bay is no longer being used replace the side and back covers and fasten with the 4 screws Adding a Pump Peristaltic pumps can only be connected to bay 1 2 and 3 Membrane pump can only be connected to bay 4 Remove the 4 screws holding side and rear covers of the pump bay Insert the pump into the rear of the Autosampler tower Once ...

Page 17: ...er using the power switch but leave the cable connected to the HI 902 Titrator 3 Insert the USB flash drive with upgrade file in the HI 921 USB slot 4 While holding down both emergency stop buttons turn on the HI 921 Autosampler Keep the buttons pressed until the status lights turn yellow The control panel should display the message Upgrading FW please wait After this message appears release the b...

Page 18: ...th Built in RFID Beakers ASTM short form glass beakers 100 150 mL HI 920 060 150 mL Plastic beakers HI 920 053 100 mL Plastic beakers Control Panel Buttons for manual operation of tray Manual operation of peristaltic or diaphragm pumps 2 line backlit display with status information Enclosure Material ABS plastic and steel Electrode Holder Material PVC Tray Material ABS plastic and acrylic Keypad M...

Page 19: ...w with RFID HI 920 11660 18 Beaker Tray 53 mm dia Single Row with RFID HI 920 11853 Plastic Beaker for HI 920 11660 20 pcs HI 920 060 Plastic Beaker for HI 920 11853 20 pcs HI 920 053 Autosampler HI 921 XY Control Panel HI 920 921 Tray Locking Screw HI 920 960 Electrode Holder HI 920 310 Temperature Sensor HI 7662 A USB Memory Stick HI 920 900 Titrant Dispensing Tube 1 5 m HI 920 280 ...

Page 20: ...astic dispensing tube for peristaltic pump HI 920 202 Tubing Set with stainless steel aspiration tube for peristaltic pump HI 920 203 Roller Tube for peristaltic pump 3 pcs HI 920 204 Membrane Pump Complete Tubing Set HI 920 212 TYGON Tube 5 m HI 920 290 Overhead Stirrer HI 920 301 Replacement Propellers 3 pcs HI 920 302 High Chemical Resistance Propellers 3 pcs HI 920 303 Cable Chain HI 920 320 2...

Page 21: ...21 INSTRUCTION MANUAL ...

Page 22: ...22 INSTRUCTION MANUAL MAN 921 11 15 ...

Reviews: