Press the ALT and TIME keys simultaneously and the display will
show the previously memorized time with the printing interval
blinking on the secondary LCD.

Any interval can be selected from among 1, 2, 5, 10, 15, 30, 60,
120 or 180 minutes by using the UP and DOWN keys.

Store the selected interval by pressing TIME once more. The hour will
blink.

To select the hour, press the UP or DOWN keys (24 hour clock).

To store the hour press the TIME key again and the minutes will
blink.

Use the UP or DOWN keys to select the correct minutes.

Press the ALT and TIME keys together to exit this mode and memorize
the minute setting.

UP

TIME

ALT

RANGE

UP

DOWN

TIME

The time, date and printing interval are now stored in the memory
even when the meter or display are switched off.

Note:

 Each time the batteries are replaced, the settings will need to

be re-entered. This can be avoided by applying an external
voltage to the unit before replacing the batteries.

VIEWING RH%, TEMPERATURE, TIME, DATE & PRINT-
ING INTERVAL WITH HI 9161

When the instrument is turned on, the dis-
play enters the TIME mode.

To view RH% on the primary display, press
the RANGE key.
To view temperature on the primary display
with decimal point, press the RANGE key.
If “----" is displayed, no probe is connected or
the probe cable is broken.

Note:

the RH% value is displayed without any symbol on the
primary display and with a "H" on the secondary LCD. The
temperature value is always displayed together with the "°C"
or "°F" symbol.

To view the current time and printing interval, press the INTV and
RANGE keys simultaneously.

To view the date, press the UP or DOWN keys.

INTV

TIME

INTV

RANGE

+

INTV

TIME

UP

DATE

DOWN

Summary of Contents for HI 91610

Page 1: ...9161C ON OFF PAPER RANGE RH 5 0 95 0 C 20 0 60 0 INTV instrum nts e UP PRINT DOWN RANGE LOG MICROPROCESSOR PRINTING THERMOHYGROMETER HI 9161 HI 91610 Printing and Logging Thermo Hygrometers w w w h a...

Page 2: ...CIFICATIONS 8 INITIAL PREPARATION 9 OPERATIONAL GUIDE 10 PRINTING RECORDING WITH HI 9161 17 PRINTING LOGGING WITH HI 91610 19 CALIBRATION 24 DATA TRANSFER TO PC HI 91610 only 27 SELF DIAGNOSTIC FUNCTI...

Page 3: ...ecording printing interval relative humidity and temperature at user selectable intervals Both RH and temperature are measured with 0 1 resolution HI 91610 combines the high accuracy and portability o...

Page 4: ...terval 6 TIME key to display current time and printing interval or to set date time printing interval with ALT 7 LOG key to start stop logging with ALT 8 LCD Liquid Crystal Display 9 PAPER key to adva...

Page 5: ...se However when in recording logging mode the meter will continue to monitor RH and temperature To reactivate the display press RANGE INITIAL PREPARATION ON OFF RANGE RANGE SPECIFICATIONS HI 9161C HI...

Page 6: ...ff until moisture has completely evaporated To accelerate the evaporation process expose the humidity sensor to a current of air For any problem in taking measurements please contact your dealer or th...

Page 7: ...ng The printing interval on the secondary LCD will blink Any interval can be selected from among 1 2 5 10 15 30 60 120 and 180 minutes by using the UP and DOWN keys Once the desired interval is select...

Page 8: ...red in the memory even when the meter or display are switched off Note Each time the batteries are replaced the settings will need to be re entered This can be avoided by applying an external voltage...

Page 9: ...YY d Printing interval in minutes e Time HH MM f RH value g Temperature value When the meter is in recording mode LOG is displayed on the secondary LCD with the RH value on the primary LCD PRINT LOG V...

Page 10: ...will print a complete set of data and the LOG symbol will appear on the secondary LCD Press the ALT and PAPER keys at the same time and the LOG symbol on display will start to blink If no key is press...

Page 11: ...DOWN TIME LOG ALT RANGE LOG TIME To reactivate the display press the TIME key Notes Once in the logging mode the interval cannot be changed Exit the logging mode press ALT and LOG together and then s...

Page 12: ...the recall mode press the LOG key Note This mode will not alter data already present in the memory PRINTING STORED DATA Once a log number is selected all or part of that logged section can be printed...

Page 13: ...Na Cl 75 4 CALIBRATION KIT 12 3 6 9 4 h CALIBRATION KIT Li Cl 11 1 Na Cl 75 4 12 3 6 9 4 h All HANNA thermo hygrometers have been precalibrated at the factory HANNA instruments uses state of the art...

Page 14: ...0 you need a 3 5 drive and a couple of minutes to follow the short instructions conveniently printed on the disk label TIME Windows is a registered Trademark of Microsoft Co Excel Copyright of Microso...

Page 15: ...s for about 1 minute Remember to reset time and date once the batteries have been reinserted ATTENTION If batteries need to be replaced and data is not to be lost power the meter through a 12 Vdc adap...

Page 16: ...er the cover preventing the pa per from advancing Replace the battery cover and tighten the screws 1 5V 1 5V 1 5V 1 5V SCREW POTS RESET CHANGING THE INK CARTRIDGE When printouts become faint it might...

Page 17: ...handle HI 762BL Liquid probe with 1 m 3 3 cable black handle HI 762BL 10 Liquid probe with 10 m 33 cable black handle HI 762L Liquid probe with 1 m 3 3 cable white handle HI 762L 10 Liquid probe with...

Page 18: ...the probe socket on the meter If the reading differs by more than 0 4 C 0 8 F from the test key value the unit is due for recalibration Choose the proper test key according to your application needs H...

Page 19: ...42 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 0...

Reviews: