23
22
Notes:
• It is recommended to use an external power supply during
logging with printing mode, especially when many printouts are
required.
• Before proceeding with logging/printing, make sure there is
enough paper for your measurements. There is no warning if the
machine runs out of paper. If this happens, data will continue to
be stored into memory, and it is always possible to print them at
a later time (see below).
• It is possible to insert a new paper roll during logging session (see
page 31).
• Once in the logging mode, the interval
cannot be changed. Exit the logging
mode first (by pressing the ALT and the
LOG keys together) and reset a new
interval.
• If the PRINT key is pressed while in
logging mode, a printout is produced
without affecting the running sample
number.
TO STOP LOGGING
Press the ALT and the LOG keys simulta-
neously. This will also generate an exit
status printout.
UP
DOWN
TIME
LOG
ALT
RANGE
UP
DOWN
TIME
LOG
ALT
RANGE
01-07-95
==02== 0005M
<>0010 *7.01
Log number
Total sample number
LOG EXIT
Date/Time
TO SCAN STORED DATA ON DISPLAY
Press the LOG key. The log number and remaining page numbers will
appear on display.
LOG
Press the ALT and LOG key simultaneously to enter the logging mode.
The current log number and the remaining page numbers will
appear for a few seconds on the display to indicate the correct
operational mode. The printer will print a complete set of data and
the "LOG" symbol will appear on the secondary LCD.
If no key is pressed, the display goes blank
after about 5 minutes and comes back on only
to print at the next logging interval. During
printing, the display shows the time, preselected
interval and the “LOG” symbol.
To reactivate the display press the TIME key.
Each printout provides the following information:
a – A running log number
b – A running sample number (in that particular log)
c – Date (DD-MM-YY)
d – Printing interval in minutes
e – Time (HH.MM)
f – RH% value
g – Temperature value
It is always possible to switch from the logging with printing function
to logging without printouts. Press ALT and PAPER at the same time
and the "LOG" symbol will start blinking to indicate that the data is
now stored into memory but no longer are printed.
UP
DOWN
TIME
LOG
ALT
RANGE
LOG
INTV
LOG
TIME
1 4 - 0 1 - 9 6
- - 0 1 - -
0 0 0 1 M
0 0 1 5
* 1 5 . 0 1
6 5 . 9 R H %
2 5 . 4 ° C
<>
a
b
c
d
e
f
g
UP
PAPER
ALT
LOG