25
TEMPERATURE COEFFICIENT
TEMPERATURE COEFFICIENT
TEMPERATURE COEFFICIENT
TEMPERATURE COEFFICIENT
TEMPERATURE COEFFICIENT
(HI 933100
(HI 933100
(HI 933100
(HI 933100
(HI 933100 and
and
and
and
and HI 9635
HI 9635
HI 9635
HI 9635
HI 9635 only
only
only
only
only)))))
SETTING THE TEMPERATURE COEFFICIENT
• Switch the meter on by pressing ON/OFF.
• Press FACTOR until the temperature coefficient is displayed.
• Change this value by pressing UP or
DOWN. This value is adjustable from 0.0
to 3.0% per degree Celsius.
This feature is particularly helpful with acids, alkalis and concentrated
salt solutions which typycally have lower values, around 1.5%/°C.
All readings the readings are standardized at 25°C.
The default temperature coefficient is 1.9%. The coefficient can be set
from 0.0% (without compensation) to 3.0% per degree Celsius. The
meter retains this coefficient even when switched off.
DETERMINATION OF THE TEMPERATURE COEFFICIENT
OF A SOLUTION
• Immerse the probe into the solution and press RANGE if necessary
to select conductivity measurements.
• Press FACTOR to display the temperature
coefficient.
• Set the temperature coefficient to "0.0%" (i.e. no compensation)
by pressing DOWN.
• Measure the conductivity of the sample at 25°C and denote it as
"C25".
µ
S
%TC
CONDUCTIVITY
FACTOR
UP
DOWN
RANGE
FACTOR
ON/OFF
DOWN
µ
S
%TC
CONDUCTIVITY