background image

1

Instruction Manual

HI 9026V

Complete Portable Kit

 for pH, mV & Temperature

Measurements in

Wine and Must

with Calibration-Check

TM

w w w . h a n n a i n s t . c o m

Summary of Contents for HI 9026V

Page 1: ...1 Instruction Manual HI 9026V Complete Portable Kit for pH mV Temperature Measurements in Wine and Must with Calibration CheckTM w w w h a n n a i n s t c o m ...

Page 2: ...The kit includes HI 9026 waterproof pH meter with Cal CheckTM HI 1048B double junction refillable pH electrode HI 7662 stainless steel temperature probe pH 3 00 and pH 7 01 buffer solutions 20 mL each HI 700635 and HI 700636 specific cleaning solutions 20 mL each HI 7082 electrode refilling solution 30 mL and syringe 100 mL plastic beaker 1 5V AA alkaline batteries 4 pcs instruction manual rugged ...

Page 3: ...tery Type 4 x 1 5V AA alkaline Battery Life Approx 500 hours of continuous use BL off Auto off User selectable 20 minutes or disabled Environment 0 to 50 C 32 to 122 F RH max 100 Dimensions meter only 196 x 80 x 60 mm 7 7 x 3 1 x 2 4 Weight meter only 500 g 1 1 lb 1 BNC plug for pH and ORP electrodes 2 Temperature probe socket 3 Liquid Crystal Display 4 RANGE key to select pH or mV 5 ON OFF key to...

Page 4: ...ions are available to condition the electrode and improve its response time see Accessories section INITIAL PREPARATION The meter is supplied complete with batteries Remove the back cover unwrap the batteries and install them while paying attention to their polarity To prepare the instrument for use connect the pH electrode and the tempera ture probe to the BNC and temperature sockets on the top o...

Page 5: ...sest full scale value If using pH electrode while in mV mode the meter will measure the mV generated by the pH electrode MEM and MR FUNCTIONS From normal measurement mode pressing the MEM key will freeze readings on display and store into internal memory the current values pH and tempera ture or mV and temperature together with condition and buffer segments The MEM tag blinks and the display remai...

Page 6: ...econdary LCD will show the second buffer to be used for calibration pH 4 01 If you re going to calibrate with a different buffer select the desired value by pressing the arrow keys Submerge the electrode approximately 4 cm 1 into the second buffer solution place the temperature probe as close as possible to the electrode and stir gently The LCD will flash NOT READY for 12 seconds then if the readi...

Page 7: ...ssue the alarm exactly 4 days after the last calibration However if at any moment the expiration value is changed e g to 7 days then the alarm will be immediately recalculated and appear 7 days after the last calibration Notes When the meter is not calibrated or calibration is cleared default values loaded there is no expired calibration and the display always shows a blinking CAL tag When an abno...

Page 8: ... 11 10 80 176 1 77 3 00 4 16 6 87 7 01 8 89 9 74 11 00 85 185 1 78 3 00 4 17 6 87 7 02 8 87 9 74 10 91 90 195 1 79 3 00 4 18 6 88 7 03 8 85 9 75 10 82 95 203 1 81 3 00 4 20 6 88 7 04 8 83 9 76 10 73 LCD MESSAGE GUIDE TAGS SYMBOLS Mode tags light up for indicating the corresponding active mode and blink for warning the user SETUP on setup menu mode has been entered CFM blinking ask confirmation of ...

Page 9: ...hour before using the electrode Note To prevent damage to the electrode remove the pH electrode from the sample before turning the meter off With the meter OFF detach the electrode from the meter and immerse it in the storage solution If batteries become weak the display will flash the battery symbol to advise the user that approximately 25 hours of working time is left It is recommended to replac...

Page 10: ...r solution 500 mL bottle HI 7007L pH 7 01 buffer solution 500 mL bottle HI 7009L pH 9 18 buffer solution 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle HI 5124 pH 12 45 buffer solution 500 mL bottle ELECTRODE MAINTENANCE SOLUTIONS HI 70000P Electrode rinsing solution 20 mL sachet 25 pcs HI 70300L Storage solution 500 mL bottle HI 7061L General cleaning solution 500 mL bottle HI 7063...

Page 11: ...3 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 765 7575 For e mail contacts and ...