Material needed
•
HI 7638
conductivity probe
•
HI 8931
conductivity controller
• Calibration solutions, according to the different models:
HI 7034
80mS/cm @25°C for
HI 8931A
HI 7030
12.88mS/cm@25°C for
HI 8931B
HI 7031
1413 µS/cm @25°C for
HI 8931C
HI 7033
84 µS/cm @25°C for
HI 8931D
PROCEDURE
• Ensure that the probe is
connected to the meter securely
by aligning the pins with the
socket, pushing the plug in and
tightening the threaded ring.
• Insert the probe into the sleeve.
• With the conductivity probe in
air, press the "MEASURE" key
to set the meter to measure-
ment mode.
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
HI 8931 WITH HI 7638
HI 8931 WITH HI 7638
HI 8931 WITH HI 7638
HI 8931 WITH HI 7638
HI 8931 WITH HI 7638