
19
Note:
•
The aspiration tube must be inserted in the titrant bottle. The dosing tip must be placed
over a rinse beaker.
•
Before starting the prime procedure make sure you are using the appropriate titrant
solution for the selected range.
• To prime the burette, select
Prime
option from
Titration
mode.
• Adjust the rinses number by pressing the and keys and press
Start
.
• The number of syringe rinses can be set between 1 and 5 (at least three rinses are recommended
to ensure that the air bubbles are completely removed).
• To pause the prime process press the
Pause
key, to continue press the
Continue
key. To stop the
prime process press the
Stop
key.